Intraoperative femoral head dislodgement during total hip arthroplasty: a report of four cases

نویسندگان

  • Ahmed Siddiqi
  • Carl T. Talmo
  • James V. Bono
چکیده

Dislodgment of trial femoral heads and migration into the pelvis during total hip arthroplasty is a rarely reported complication with limited published cases. There are three primary mechanisms of femoral head separation: dislodgement during reduction attempt, disassociation from anterior dislocation while assessing anterior stability, and during dislocation after implant trialing. If the trial femoral migrates beyond the pelvic brim, it is safer to finish the total hip arthroplasty and address the retained object after repositioning or in a planned second procedure with a general surgeon. We recommend operative retrieval since long-term complications from retention or clinical results are lacking.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intraoperative Migration of the Trial Femoral Head into the Pelvis during Total Hip Arthroplasty –report of Two Cases–

Various complications occur during total hip arthroplasty (THA). It is common to implement trial treatments of various implants during THA. Here we report on 2 cases in which the trial femoral head fell into the pelvis and migrated during THA. The trial head was not retrieved but left inside in one case, while the other presented the opportunity for a possible retrieval. It is important to reco...

متن کامل

Intrapelvic Dislocation of a Femoral Trial Head During Primary Total Hip Arthroplasty Requiring Laparotomy for Retrieval

BACKGROUND AND PURPOSE Total hip arthroplasty (THA) is a safe and reliable surgical procedure. However, THA also has intra- and postoperative complications. A dreaded and frustrating intraoperative complication during total hip arthroplasty is dislocation of the femoral trial head from the neck into the pelvis. METHODS Here, we report on the case of a 71-year old female patient with osteoarth...

متن کامل

Fatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature

Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...

متن کامل

Late Disassembly of Femoral Head and Neck of A Modular Primary Total Hip Arthroplasty

INTRODUCTION Modular total hip arthroplasty system are now widely used, as these components increase the flexibility during primary and revision total hip arthoplasty. But this modularity itself associated with some risk of intraoperative and postoperative complications. CASE REPORT We report a case of late disassembly of a primary total arthroplasty in a 42 years old patient five years after...

متن کامل

Intraoperative dislocation of the trial bipolar cup into the pelvis during bipolar hip arthroplasty – A case report☆

INTRODUCTION Bipolar hip arthroplasty is a good option for treating femoral neck fractures, although some contraindications have been indicated. We report a case of intraoperative dislocation of the trial bipolar cup into the pelvis during bipolar hip arthroplasty. CASE PRESENTATION A 74-year-old woman underwent bipolar hip arthroplasty for a femoral neck fracture (AO31-B2). She was placed in...

متن کامل