A case of Durata ICD lead coil externalization: Inside-out lead abrasion?
نویسندگان
چکیده
منابع مشابه
A case of Durata ICD lead coil externalization: Inside-out lead abrasion?
The St Jude Medical Durata lead was produced in response to the recalled Riata and Riata ST leads in November 2011 after concerns of lead durability and inside-out abrasion. In order to protect against this susceptibility, the Riata ST Optim and Durata leads were coated with an abrasionresistant layer of silicone–polyurethane copolymer (Optim) tubing. We report a case of Durata lead extraction ...
متن کاملUnusual fracture in a Durata lead with shock coil fragmentation and cable externalization
From the *Division of Advanced Arrhythmia Management, the Second Department of Internal Medicine, Kyorin University School of Medicine, Mitaka, Japan, Division of Cardiology, the Second Department of Internal Medicine, Kyorin University School of Medicine, Mitaka, Japan, and Clinical Research Division for Heart Rhythm Management, Department of Cardiology, Tokyo Women’s Medical University, Tokyo...
متن کاملTo the editor - Inside-out insulation failure of a defibrillator lead with abrasion resistant-coating.
Clinical report A 78-year-old woman had a cardiac resynchronization implantable cardioverter-defibrillator implanted in January 2008 for nonischemic cardiomyopathy with New York Heart Association class III heart failure, left ventricular ejection fraction 22%, and left bundle branch block. A St Jude Medical Model 7121 Durata right ventricular lead (St Jude Medical, St Paul, MN) and a Medtronic ...
متن کاملSingle-coil Versus Dual-coil ICD Lead Shock Efficacy in a Large ICD Registry
Single-coil leads account for 5–10% of transvenous implantable cardioverterdefibrillator (ICD) leads. Single-coil leads improve extractability while allowing adequate safety margin for defibrillation in new generation devices (output of 35–41 joules). Limited clinical shock efficacy data exist in patients having single-coil ICD systems. A retrospective analysis was conducted from St. Jude Medic...
متن کاملICD lead abandonment is without risk? A case of “lead on lead crime”
Introduction Lead failure rates of implantable cardioverter-defibrillators (ICDs) vary depending on design and composition, and can be as high as 40% a few years post implant. The Sprint Fidelis lead (Medtronic, Minneapolis, MN) was recalled in 2007 because it is prone to fracture, particularly to its pacesense conductor, which predisposes patients to inappropriate shocks; its yearly failure ra...