A fractured and immovable ceramic liner in a screw-fixed acetabular shell during total hip arthroplasty: a case report.

We report a complication during total hip arthroplasty using an uncemented press-fit acetabular shell and ceramic liner for a woman with osteoarthritis. Fracture of the ceramic liner occurred during its insertion owing to non-concentric reduction. The main body of the ceramic liner became wedged in the acetabular shell and was resistant to removal. The shell itself was immovable owing to the pr...

متن کامل

Extraction of a well fixed but fractured ceramic acetabular liner

Ceramic fractures have been reported to occur in hip replacements, but the techniques used to extract a well fixed but fractured component are not commonly described. We present a case of ceramic liner fracture and validate a modification of a previously reported extraction technique that allowed us to save the acetabular cup and insert a polyethylene liner. With an increasing trend in ceramic ...

متن کامل

Catastrophic Failure of the Acetabular Polyethylene Liner in Ceramic-on-Polyethylene Total Hip Arthroplasty

INTRODUCTION Catastrophic polyethylene failure is a rare complication of ceramic-on-polyethylene total hip arthroplasty due to the favorable tribological characteristics of ceramic. Failure of the polyethylene liner can be disastrous, increasing periprosthetic osteolysis, metallosis, and risk of dislocation. Complications associated with ceramic-on-polyethylene articulations have been studied e...

متن کامل

Metallosis after Exchange of the Femoral Head and Liner following Ceramic Acetabular Liner Dissociation in Total Hip Arthroplasty with a Modular Layered Acetabular Component

The type of bearing material that should be used in revision surgery after the failure of ceramic-on-ceramic total hip arthroplasty (THA) remains controversial. In the case of ceramic fracture, the residual ceramic particles can cause consequent metallosis when metal implants are used for revision THA. On the other hand, in the case of THA failure without ceramic fracture, revision THA with a m...

متن کامل

Fracture of Ceramic Liner and Head in a Total Hip Arthroplasty with a Sandwich Type Cup

Due to its advantages, ceramic-on-ceramic bearings have been widely used in young patients for almost 30 years. Long-term survivorship, low wear, and low biological reactivity to particles are some of its characteristics. Even though this material has had a lot of improvements, the risk of fracture is one of the concerns. There have been reports of fracture of ceramic in the acetabular liner an...