PURPOSE Brain single photon emission computed tomography (SPECT) imaging is an important clinical tool, with unique tracers for studying neurological diseases. Nowadays, most commercial SPECT systems are combined with x-ray computed tomography (CT) in so-called SPECT/CT systems to obtain an anatomical background for the functional information. However, while CT images have a high spatial resolu...

متن کامل

Collimator design for a clinical brain SPECT/MRI insert

The Institute of Nuclear Medicine, University College London, London, UK This project’s goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging. We assume the stationary SPECT insert will consist of two rings of ∼5x5-cm SiPM-based detectors insensitive to magnetic fields, with 0.8-mm intrinsic resolution. The maximum diameter is 44.5 cm, the minimum di...

متن کامل

Performance of a novel collimator for high-sensitivity brain SPECT.

We assessed improvements in performance in detection and estimation tasks due to a novel brain single photon computed tomography collimator. Data were acquired on the CeraSPECT scanner using both new and standard collimators. The new variable focusing collimator SensOgrade samples the projections unequally, with central regions more heavily represented, to compensate for attenuation of counts f...

متن کامل

Evaluation of a partial ring design for the INSERT SPECT/MRI system.

Institute of Nuclear Medicine, University College London, UK The aim of the INSERT project is to develop a SPECT insert for a commercial MRI system, for performing simultaneous SPECT/MRI brain studies in humans. We have previously investigated various design options for the detector system, based on a complete ring of detectors. We are now considering a partial ring, due to space limitations. W...

متن کامل

A novel approach to multipinhole SPECT for myocardial perfusion imaging.

UNLABELLED Myocardial perfusion imaging with SPECT remains critically important for diagnosing, assessing, and evaluating treatment of coronary artery disease. However, conventional rotational SPECT suffers from prolonged study times because of relatively low detection efficiency. We therefore have investigated a multipinhole collimator that could improve the detection efficiency in cardiac SPE...

متن کامل