Comparison of two cryogenic radiometers at NIST
نویسندگان
چکیده
منابع مشابه
Comparison of Two Cryogenic Radiometers at NIST
Two cryogenic radiometers from NIST, one from the Optical Technology Division and the other from the Optoelectronics Division, were compared at three visible laser wavelengths. For this comparison, each radiometer calibrated two photodiode trap detectors for spectral responsivity. The calibration values for the two trap detectors agreed within the expanded (k = 2) uncertainties. This paper desc...
متن کاملIntercomparison of the LBIR Absolute Cryogenic Radiometers to the NIST Optical Power Measurement Standard
The Low Background Infrared calibration (LBIR) facility at the National Institute of Standards and Technology (NIST) presently maintains four absolute cryogenic radiometers (ACRs) which serve as standard reference detectors for infrared calibrations performed by the facility. The primary standard for optical power measurements at NIST-Gaithersburg has been the High Accuracy Cryogenic Radiometer...
متن کاملcomparison of the amount of debris extruded apically in two rotary techniques: flexmaster and m2
چکیده ندارد.
15 صفحه اولThe Use of Cryogenic HEMT Amplifiers in Wide Band Radiometers
Advances in device fabrication, modelling and design techniques have made wide band, low noise cryogenic amplifiers available at frequencies up to 106 GHz. Microwave radiometry applications as used in radio astronomy capitalize on the low noise and large bandwidths of these amplifiers. Radiometers must be carefully designed so as to preclude sensitivity degradations caused by small, low frequen...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Journal of Research of the National Institute of Standards and Technology
سال: 2001
ISSN: 1044-677X
DOI: 10.6028/jres.106.029