Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
نویسندگان
چکیده
منابع مشابه
Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning
This work considers a previously overlooked uncertainty present in film dosimetry which results from moderate curvature of films during the scanning process. Small film samples are particularly susceptible to film curling which may be undetected or deemed insignificant. In this study, we consider test cases with controlled induced curvature of film and with film raised horizontally above the sc...
متن کاملRadiochromic Film Dosimetry
Published by the American Association of Physicists in Medicine DISCLAIMER: This publication is based on sources and information believed to be reliable, but the AAPM and the editors disclaim any warranty or liability based on or relating to the contents of this publication. The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as...
متن کاملReference dosimetry using radiochromic film
The objectives of this study are to identify and quantify factors that influence radiochromic film dose response and to determine whether such films are suitable for reference dosimetry. The influence of several parameters that may introduce systematic dose errors when performing reference dose measurements were investigated. The effect of the film storage temperature was determined by comparin...
متن کاملRadiochromic film dosimetry in water phantoms.
Radiochromic film is investigated for use in dosimetry in water phantoms as opposed to solid phantoms. Investigations are performed to measure the penetration rates of water into radiochromic film and to assess the effects on optical density that this penetration causes. The effects of film orientation during irradiation in water are also tested. Results show that only a small penetration rate ...
متن کاملDosimetry of blood irradiation with radiochromic film.
It has been shown that radiochromic film is an ideal dosimeter for assessment and verification of delivered dose to irradiated blood products. Using a parallel opposing two-field technique on a medical linear accelerator, blood is irradiated to diminish the risk of transfusion-associated graft vs. host disease (TA-GVHD). The blood products are irradiated in a Perspex blood box to an applied dos...