Failure mode and effect analysis-based quality assurance for dynamic MLC tracking systems
نویسندگان
چکیده
منابع مشابه
Failure mode and effect analysis-based quality assurance for dynamic MLC tracking systems.
PURPOSE To develop and implement a failure mode and effect analysis (FMEA)-based commissioning and quality assurance framework for dynamic multileaf collimator (DMLC) tumor tracking systems. METHODS A systematic failure mode and effect analysis was performed for a prototype real-time tumor tracking system that uses implanted electromagnetic transponders for tumor position monitoring and a DML...
متن کاملDetailed analysis of latencies in image-based dynamic MLC tracking.
PURPOSE Previous measurements of the accuracy of image-based real-time dynamic multileaf collimator (DMLC) tracking show that the major contributor to errors is latency, i.e., the delay between target motion and MLC response. Therefore the purpose of this work was to develop a method for detailed analysis of latency contributions during image-based DMLC tracking. METHODS A prototype DMLC trac...
متن کاملFailure Mode and Effect Analysis Power Plant Boiler
The current electricity demand is increasing, and now the government has involved third parties in the implementation of electricity so that investors compete in building infrastructure in order to apply electricity. Thermal power is one source that has a fast break event point compared to other resources that more interested investors even with all forms of pollution caused. A form of heat pow...
متن کاملDesign Software Failure Mode and Effect Analysis using Fuzzy TOPSIS Based on Fuzzy Entropy
One of the key pillars of any operating system is its proper software performance. Software failure can have dangerous effects and consequences and can lead to adverse and undesirable events in the design or use phases. The goal of this study is to identify and evaluate the most significant software risks based on the FMEA indices with respect to reduce the risk level by means of experts’ opini...
متن کاملDynamic MLC Tracking Using 4D Lung Tumor Motion Modelling and EPID Feedback
Background: Respiratory motion causes thoracic movement and reduces targeting accuracy in radiotherapy. Objective: This study proposes an approach to generate a model to track lung tumor motion by controlling dynamic multi-leaf collimators. Material and Methods: All slices which contained tumor were contoured in the 4D-CT images for...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Medical Physics
سال: 2010
ISSN: 0094-2405
DOI: 10.1118/1.3517837