Novel Size-Variable Dedicated Rodent Oxygenator for ECLS Animal Models—Introduction of the “RatOx” Oxygenator and Preliminary In Vitro Results

نویسندگان

چکیده

The overall survival rate of extracorporeal life support (ECLS) remains at 60%. Research and development has been slow, in part due to the lack sophisticated experimental models. This publication introduces a dedicated rodent oxygenator (“RatOx”) presents preliminary vitro classification tests. RatOx an adaptable fiber module size for various Gas transfer performances over different blood flows sizes were tested according DIN EN ISO 7199. At maximum possible amount effective surface area flow 100 mL/min, performance was 6.27 mL O2/min 8.2 CO2/min, respectively. priming volume largest is 5.4 mL, while smallest configuration with single mat layer 1.1 mL. novel ECLS system evaluated demonstrated high degree compliance all pre-defined functional criteria rodent-sized animal We intend become standard testing platform scientific studies on therapy technology.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Heart pump and oxygenator.

1124 T ms PAPER is a general review of the development of the artificial heart-lung to facilitate open-heart surgery. At the close of World War II many centers began investigating the possibility of total cardiac bypass. Oven the past decade, pump oxygenerators of various types have become popular and recent clinical successes throughout the world have given further impetus to the study of prob...

متن کامل

The transition from the bubble oxygenator to the microporous membrane oxygenator.

The transition from bubble oxygenators to microporous membrane oxygenators is a good example of how industry working with customers and material suppliers can help improve a surgical technique, such as cardiopulmonary bypass. The development is characterized by many individuals who, through their inventiveness and determination, made the transition possible. While the medical personnel involved...

متن کامل

Predicting oxygenator clinical performance from laboratory in-vitro testing.

Knowledge and predictability of oxygenator performance is vital to safe and effective conduct of cardiopulmonary bypass. The determination of oxygenator performance in the laboratory, however, is carried out under a strict set of conditions established by the Association for the Advancement of Medical Instrumentation (AAMI). This performance data is of limited value in the clinical setting wher...

متن کامل

Evaluation of infant membrane oxygenator in sheep.

OBJECTIVE To analyze the security and efficacy of a new membrane oxygenator, the so-called OXM - 1500. METHODS From May 2005 to September 2006, six sheep of Santa Inês breed (five male and one female, respectively) were studied. The average body weight was 14.1 (+/-5) kg, body surface 0.6 (+/-0.2) m(2) and a mean age 3.8 (+/-1.5) months. All of them were submitted to extracorporeal circulatio...

متن کامل

Fentanyl uptake by the scimed membrane oxygenator.

With the initiation of cardiopulmonary bypass (CPB), using a membrane oxygenator, the drop in circulating fentanyl concentration is greater than can be attributed to dilution alone. This study examined the Scimed brand (2A-800) membrane oxygenator as a site of fentanyl binding. Initial experiments used an assembled CPB circuit. Subsequent dissection and analysis of the oxygenator revealed that ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ژورنال

عنوان ژورنال: Micromachines

سال: 2023

ISSN: ['2072-666X']

DOI: https://doi.org/10.3390/mi14040800