COBALT PHTHALOCYANINE MODIFIED CARBON PASTE ELECTRODE AS A POTENTIOMETRIC SENSOR FOR DETERMINATION OF ASCORBIC ACID
A carbon Paste electrode modified with cobalt (II) phthalocyanine (CoPC) was used as a sensitive potentiometric sensor for determination of ascorbic acid in aqueous solutions. The potential response of the electrode is a linear function of ascorbic acid concentration over the range 5 10 to 10 M (?1- 1800 µg ML ) with a Nernstian slope of 56 mV at pH 7. The limit of detection was 0.2 µg ...
متن کاملcobalt phthalocyanine modified carbon paste electrode as a potentiometric sensor for determination of ascorbic acid
a carbon paste electrode modified with cobalt (ii) phthalocyanine (copc) was used as a sensitive potentiometric sensor for determination of ascorbic acid in aqueous solutions. the potential response of the electrode is a linear function of ascorbic acid concentration over the range 5 10 to 10 m (?1- 1800 µg ml ) with a nernstian slope of 56 mv at ph 7. the limit of detection was 0.2 µg ml ascor...
متن کاملFlow-Injection Amperometric Determination of Ascorbic Acid Using a Graphite-Epoxy Composite Electrode Modified with Cobalt Phthalocyanine
A flow injection method is reported for the determination of ascorbic acid based on amperometric detection, using a cobalt(II) phthalocyanine (CoPc) modified graphite-epoxy electrode. The amperometric respons was evaluated with regard to pH, ionic strength of the electrolyte, flow rate of the carrier solution, injected sample volume and conditioning time of the electrode. The limit of detec...
متن کاملflow-injection amperometric determination of ascorbic acid using a graphite-epoxy composite electrode modified with cobalt phthalocyanine
a flow injection method is reported for the determination of ascorbic acid based on amperometric detection, using a cobalt(ii) phthalocyanine (copc) modified graphite-epoxy electrode. the amperometric respons was evaluated with regard to ph, ionic strength of the electrolyte, flow rate of the carrier solution, injected sample volume and conditioning time of the electrode. the limit of detection...
متن کاملFlow-injection electrochemical determination of citric acid using a cobalt(II)-phthalocyanine modified carbon paste electrode.
The electrocatalytic activity of cobalt(II)-phthalocyanine immobilized in a carbon paste electrode for citric acid oxidation under a diffusion-controlled electrochemical process was employed to develop a selective flow injection amperomettric analysis method. Under optimized conditions, rapid and reliable determinations of citric acid (65 samples per hour) were achieved for analyses of commerci...