Study on virtual zero power of maglev left ventricular assist device

نویسندگان

چکیده

A virtual zero power (VZP) method was used to reduce the consumption of hybrid magnetic bearings (HMBs) in left ventricular assist devices (LVADs). The mathematics HMB system LVAD established, and stability analyzed with a proportional-derivative (PD) controller. unbalanced forces on rotor can be identified by output voltage PD controller, together negative stiffness permanent magnet, suspended at point where rotor’s reluctant force is minimum lower consumption. Experiments were conducted measure bearing different rotational speeds, trajectories demonstrated. results showed using VZP consumed 0.28 W averagely 1000-4000 rpm LVAD.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improvement of Left Ventricular Assist Device (LVAD) in Artificial Heart Using Particle Swarm Optimization

In this approach, the Left ventricular assist pump for patients with left ventricular failure isused. The failure of the left ventricle is the most common heart disease during these days. Inthis article, a State feedback controller method is used to optimize the efficiency of a samplingpump current. Particle Swarm Algorithm, which is a set of rules to update the position andvelocity, is applied...

متن کامل

Computational fluid dynamics analysis of a maglev centrifugal left ventricular assist device.

The fluid dynamics of the Thoratec HeartMate III (Thoratec Corp., Pleasanton, CA, U.S.A.) left ventricular assist device are analyzed over a range of physiological operating conditions. The HeartMate III is a centrifugal flow pump with a magnetically suspended rotor. The complete pump was analyzed using computational fluid dynamics (CFD) analysis and experimental particle imaging flow visualiza...

متن کامل

Design an Equivalent Left Ventricular Assist Device for Medical Equipment Labs

LVAD is a mechanical pump supporting a weak heart function and blood flow. Sometimes, the heart may not recover fast enough to take over the pumping action immediately after surgery, in such patients a temporary support device has been employed to maintain the pumping action until the patient’s own heart recovers. This device can be considered as a temporary alternative before the process of ar...

متن کامل

Left ventricular-assist device arrest: total thrombosis of a continuous flow left ventricular-assist device.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . doi:10.1093/ehjci/jev290 Online publish-ahead-of-pr...