Abstract Background Haemodynamic assessment in and before pregnancy is becoming increasingly important relation to complications outcomes. Different methodologies exist but there no gold-standard technique for non-invasive measurement of cardiac output (CO). We sought assess two methods CO healthy women undergoing vitro fertilisation cycles (IVF). This was a prospective longitudinal study 71 aged 18–44 years planning IVF measurements obtained via inert gas rebreathing (IGR) using Innocor ™ whole-body bio-impedance (WBI) Nicas the reproducibility between methods. Four visits occurred at which both techniques were used: initial assessment, embryo transfer, day test 4 weeks post-transfer (regardless whether conception occurred). Cross-sectional agreement assessed calculation bias, percentage error limits (LOA) Bland–Altman analysis. Longitudinal 4-quadrant plot with concordance rate, angular bias radial (%). Results One hundred thirteen from 44 participants suitable cross-sectional (Bland–Altman) IGR (Innocor ) Mean 4.61 L/min 5.05 WBI (Nicas ). The 0.44 L/min. 76% intra-correlation coefficient 0.135 (95% CI −0.43–0.306). Fifty-nine 28 (4Q-plot) rate 64.4%, – 0.14, + − 13.25°. Conclusion There poor techniques. These cannot be used interchangeably when measuring IVF, these results may more generalizable, peri-conception period.
منابع مشابه
Non-invasive Assessment of Cardiac Output Using an Inert Gas Rebreathing Device: Repeatability and Reproducibility in Normal Controls
P134 Table Pre-CDM Post-CDM Z score p Value* 3 Months 0.60 (0–6) 0.22 (0–4) 25.58 ,0.001 6 Months 0.86 (0–10) 0.44 (0–5) 25.23 ,0.001 12 Months 1.14 (0–13) 0.79 (0–9) 22.82 ,0.005 *Wilcoxon rank. CDM, chronic disease management. Poster presentations Thorax 2008;63(Suppl VII):A74–A160 A127 group.bmj.com on October 20, 2017 Published by http://thorax.bmj.com/ Downloaded from
متن کاملNon-invasive measurement of cardiac output in heart failure patients using a new foreign gas rebreathing technique.
Values of effective pulmonary blood flow (Q(EP)) and cardiac output, determined by a non-invasive foreign gas rebreathing method (CO(RB)) using a new infrared photoacoustic gas analysing system, were compared with measurements of cardiac output obtained by the direct Fick (CO(FICK)) and thermodilution (CO(TD)) methods in patients with heart failure or pulmonary hypertension. In 11 patients, of ...
متن کاملThe reliability of noninvasive cardiac output measurement using the inert gas rebreathing method in patients with advanced heart failure.
BACKGROUND Impaired cardiac output (CO) is a key element of heart failure (HF). So far, there has been no simple, reliable, inexpensive and non-invasive CO measurement method feasible for clinical practice. Not a single diagnostic test has been elaborated to diagnose and monitor HF. The aim of the study was the evaluation of the reliability of a new, non-invasive CO measurement device utilizing...
متن کاملNon-invasive measurement of cardiac output by whole-body bio-impedance during dobutamine stress echocardiography: clinical implications in patients with left ventricular dysfunction and ischaemia.
OBJECTIVES To compare non-invasive determination of cardiac index (CI) by whole body electrical bioimpedance using the NICaS apparatus and Doppler echocardiography, and the role of cardiac power index (Cpi) and total peripheral resistance index (TPRi) calculation during dobutamine stress echocardiography (DSE). SUBJECTS AND METHODS We enrolled 60 consecutive patients undergoing DSE. Patients ...
متن کاملValidation of Noninvasive Measurement of Cardiac Output Using Inert Gas Rebreathing in a Cohort of Patients With Heart Failure and Reduced Ejection Fraction.
BACKGROUND Cardiac output (CO) is a key indicator of cardiac function in patients with heart failure. No completely accurate method is available for measuring CO in all patients. The objective of this study was to validate CO measurement using the inert gas rebreathing (IGR) method against other noninvasive and invasive methods of CO quantification in a cohort of patients with heart failure and...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Artery Research
سال: 2022
ISSN: ['1876-4401', '1872-9312']
DOI: https://doi.org/10.1007/s44200-022-00019-9