accuracy of linear measurements using cone beam computed tomography in comparison with clinical measurements

نویسندگان

amir reza rokn rokn

kazem hashemi

solmaz akbari

mohammad javad kharazifard

چکیده

objectives: this study sought to evaluate the accuracy and errors of linear measurements of mesiodistal dimensions of kennedy class iii edentulous space using cone beam computed tomography (cbct) in comparison with clinical measurements. materials and methods: nineteen kennedy class iii dental arches were evaluated. an impression was made of each dental arch and poured with dental stone. the distance was measured on dental cast using a digital vernier caliper with an accuracy of 0.1mm and on cbct scans. finally, the linear mesiodistal measurements were compared and the accuracy of cbct technique was evaluated by calculating absolute value of errors, intra-class correlation coefficient and simple linear regression model. results: in comparison with the cast method, estimation of size on cbct scans had an error of -8.46% (underestimation) to 5.21% (overestimation). in 26.5% of the cases, an accepted error of ±1% was found. the absolute value of errors was found to be in the range of 0.21-8.46mm with an average value of 2.86 ±2.30mm. conclusions: although the measurements revealed statistically significant differences, this does not indicate a lower accuracy for the cbct technique. in fact, cbct can provide some information as a paraclinical tool and the clinician can combine these data with clinical data and achieve greater accuracy. undoubtedly, calibration of data collected by clinical and paraclinical techniques and the clinician’s expertise in use of cbct software programs can increase the accuracy of implant placement.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of Linear Measurements Using Cone Beam Computed Tomography in Comparison with Clinical Measurements

OBJECTIVES This study sought to evaluate the accuracy and errors of linear measurements of mesiodistal dimensions of Kennedy Class III edentulous space using cone beam computed tomography (CBCT) in comparison with clinical measurements. MATERIALS AND METHODS Nineteen Kennedy Class III dental arches were evaluated. An impression was made of each dental arch and poured with dental stone. The di...

متن کامل

The Accuracy of Cone-beam Computerized Tomography Linear Measurements in Human Dry Mandible

Background & Aims: The aim of this study was to evaluate the accuracy of Cone-beam Computerized Tomography (CBCT) linear measurements in dry mandibles by comparing them with direct measurements. Methods: Two human dry mandibles were used in this study. Thirteen sites were selected on each mandible. Specimens were scanned by CBCT and linear measurements were made using Planmeca Romexis 2.3.0.R s...

متن کامل

Effect of Slice Thickness on the Accuracy of Linear Measurements Made on Cone Beam Computed Tomography Images (InVitro

Objective: Cone beam computed tomography (CBCT) is applied for the imaging of the maxillofacial and dental structures, particularly for surgical treatments and dental implants. The aim of this study was to assess the effect of slice thickness on the accuracy of CBCT linear measurements. Methods: In this diagnostic accuracy study, forty-two titanium pins with the same dimensions were inserted...

متن کامل

accuracy and reliability of linear measurements using tangential projection and cone beam computed tomography

background: providing a cross-sectional image is essential for preimplant assessments. computed tomography (ct) and cone beam ct (cbct) images are very expensive and provide high radiation dose. tangential projection is a very simple, available, and low-dose technique that can be used in the anterior portion of mandible. the purpose of this study was to evaluate the accuracy of tangential proje...

متن کامل

Accuracy of Linear Measurements of Galileos Cone Beam Computed Tomography in Normal and Different Head Positions

Introduction. The aim of this study was to determine the accuracy of linear measurements in dry human skulls in ideal position and different deviated positions of the skull. Methods. 6 dry human skulls were included in the study. Opaque markers were attached to alveolar bone. Buccolingual and mesiodistal distances and heights were measured in 5 different regions of either jaws using a digital c...

متن کامل

Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCT-NewTom).

OBJECTIVES The conical beam computed tomography (CBCT) technique presents an innovation of tomographic imaging systems and subsequent volumetric image reconstruction for dentistry. When compared with other methods of tomographic imaging CBCT is characterized by rapid volumetric image acquisition from a single low radiation dose scan of the patient. The NewTom (NewTom 9000; Quantitative Radiolog...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
journal of dentistry, tehran university of medical sciences

جلد ۱۳، شماره ۵، صفحات ۳۳۳-۳۳۹

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023