malfunction of totally implantable central venous ports

نویسندگان

hyo-cheol kim department of radiology, seoul national university hospital, seoul, korea; department of radiology, seoul national university hospital, seoul, korea. tel: 82-220722584, fax: 82-27437418

saebeom hur department of radiology, seoul national university hospital, seoul, korea

hoyong jeon department of radiology, seoul national university hospital, seoul, korea

چکیده

conclusion our data suggest that catheter malfunctions of ticvps are more common in females than males. the incidence of ticvp malfunctions does not differ between the two devices (celsite® vs. humanport®). background malfunctions of totally implantable central venous ports (ticvps) have become a problem, as the usage of ticvps has increased enormously. objectives this study evaluated factors related with catheter malfunctions of ticvps. patients and methods between january 2010 and june 2012, 1,740 ticvps in 1,740 patients (874 men and 866 women) with an average age of 57.7 ± 12.8 years (range: 15 - 91 years) were implanted by an interventional radiology team at our institute. catheter malfunctions were retrospectively analyzed. in the prospective study, we randomly allocated 176 patients to two kinds of ticvps in a 1:1 assignment ratio. the primary outcome was the malfunction of ticvps. results in the retrospective study, the 32 malfunctioning ticvps were caused by fibrin sheath formations (n = 15), chamber thrombosis (n = 8), ticvp rotation (n = 5), catheter migration (n = 2), and blood clots within the catheters (n = 2). multivariate analysis showed that being female was a significant factor for poor patency rates of ticvps (hazard ratio: 5.06; 95% confidence interval (ci) 1.32 - 19.46, p = 0.018). in the prospective study, two chamber thromboses occurred in celsite® (n = 1) and humanport® (n = 1). the primary patency rates of both celsite® and humanport® were 98.9% at 6 months, respectively.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Insertion of Totally Implantable Central Venous Access Devices by Surgeons

PURPOSE The aim of this study is to evaluate the results for the insertion of totally implantable central venous access devices (TICVADs) by surgeons. METHODS Total 397 patients, in whom TICVADs had been inserted for intravenous chemotherapy between September 2008 and June 2014, were pooled. This procedure was performed under local anesthesia in an operation room. The insertion site for the T...

متن کامل

The late complications of totally implantable central venous access ports: the results from an Italian multicenter prospective observation study.

PURPOSE The principal aim of this study is to analyze the incidence of late complications in oncologic patients with totally implanted central venous access ports. METHODS A prospective multicenter observational study was conducted in 26 Italian oncologic outpatient clinics. 1076 cancer patients with Totally Implanted Central Venous Access Ports (TIAP) were observed. 515 devices were observed...

متن کامل

A new approach to mitigate biofilm formation on totally implantable venous access ports.

Use of intravascular catheters for patient care may be associated with increased risk of central line–associated bloodstream infections. It is estimated that up to 18 000 such infections occurred in intensive care units and up to 23 000 in inpatient wards of acute care hospitals in the United States in 2009 [1]. These device-associated infections result in significant morbidity, mortality, and ...

متن کامل

A Retrospective Clinical Study: Complications of Totally Implanted Central Venous Access Ports

BACKGROUND When managing patients who require repeated venous access, gaining a viable intravenous route has been problematic. To improve the situation, various studies on techniques for venous access have been conducted. The aim of this study is to evaluate the clinical results of complications following totally implanted central venous access port (TICVAP) insertion. METHODS A retrospective...

متن کامل

Safety of a Totally Implantable Central Venous Port System with Percutaneous Subclavian Vein Access

BACKGROUND The role of totally implantable central venous port (TICVP) system is increasing. Implantation performed by radiologist with ultrasound-guided access of vein and fluoroscope-guided positioning of catheter is widely accepted nowadays. In this article, we summarized our experience of TICVP system by surgeon and present the success and complication rate of this surgical method. MATERI...

متن کامل