loss of guide wire: a rare complication of intra-aortic balloon pump insertion

نویسندگان

manouchehr hekmat shahid modarres hospital, shahid beheshti university of medical sciences, tehran, iran.

hamid ghaderi shahid modarres hospital, shahid beheshti university of medical sciences, tehran, iran.

seyedehadeleh mirjafari brain and spinal injury research center (basir), tehran university of medical sciences, tehran, iran.

چکیده

at the final stages of a coronary artery bypass graft operation on a 64-year-old man, an experienced physician attempted to insert an intra-aortic balloon pump into the femoral artery via the seldinger technique. however, while the balloon pump was being passed over the guide wire, the latter was completely lost. the guide wire should be held at the tip at all times to prevent passage into the vessel. strict adherence to this rule will prevent guide-wire loss, which is, albeit rare and completely avoidable, a potentially life-threatening complication of central vein or artery catheterization, with reported fatality rates of up to 20% when the whole wire is lost. the literature contains several reports on guide-wire loss during central venous, arterial, and hemodialysis catheterization, but we report for the first time the loss of a guide wire as a rare complication of intra-aortic balloon pump insertion.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Loss of Guide Wire: A Rare Complication of Intra-Aortic Balloon Pump Insertion

At the final stages of a coronary artery bypass graft operation on a 64-year-old man, an experienced physician attempted to insert an intra-aortic balloon pump into the femoral artery via the Seldinger technique. However, while the balloon pump was being passed over the guide wire, the latter was completely lost. The guide wire should be held at the tip at all times to prevent passage into the ...

متن کامل

The Evaluation of Survival in Patients Who Need Intra Aortic Balloon Pump (IABP) after Cardiac Surgery

Introduction: Insome cases, assist devices are required to ensure an adequate cardiac output after cardiac surgery. Intra-aortic balloon pump (IABP) is the most commonly used cardiac assist device, which provides first-line support for patients with heart failure. The aim of this study was to determine factors affecting the mortality rate of patients receiving IABP. Materials & Methods:  In thi...

متن کامل

Complications of Intra - aortic Balloon Insertion and

One hundred consecutive patients considered to be candidates for intra-aortic balloon pumping (IABP) are reviewed from a standpoint of complications and difficulties encountered with balloon catheter insertion. A single attempt at positioning the balloon in the descending thoracic aorta proved successful in 73% of these patients, but the standard femoral artery approach proved difficult in 27% ...

متن کامل

Intra-Aortic Balloon Pump Rupture and Entrapment

Intra-aortic balloon pump is used frequently to support a failing myocardium in cardiac patients. Due to the invasive nature of this device, usage is accompanied by consistent risk of complications. Balloon rupture, although it occurs rarely, may lead to entrapment if diagnosis delays. A 78-year male who underwent cardiac surgery experienced balloon rupture and entrapment in the right femoral a...

متن کامل

Intra-aortic balloon pump for oesophageal dilatation.

We read with interest the article by Loubani et al. [1] regarding the use of intra-aortic balloon pumps (IABP) for dilating oesophageal strictures. There are fundamental reasons why IABP is not the instrument of choice for oesophageal dilatation, even in difficult cases. IABP was never designed as a dilator. Its function is to produce diastolic augmentation of coronary and systemic blood flow b...

متن کامل

Predictors of Intra-Aortic Balloon Pump Insertion in Coronary Surgery and Mid-Term Results

BACKGROUND We aimed to investigate the preoperative, operative, and postoperative factors affecting intra-aortic balloon pump (IABP) insertion in patients undergoing isolated on-pump coronary artery bypass grafting (CABG). We also investigated factors affecting morbidity, mortality, and survival in patients with IABP support. METHODS Between January 2002 and December 2009, 1,657 patients unde...