transcatheter atrial septal defect closure using occlutech figulla device: a two-center experience

نویسندگان

hojat mortezaeian rajaie cardiovascular, medical and research center, iran university of medical sciences, tehran, iran.

keyhan sayadpour-zanjani children’s medical center (pediatrics center of excellence), tehran university of medical sciences, tehran, iran.

elaheh malakan-rad children’s medical center (pediatrics center of excellence), tehran university of medical sciences, tehran, iran.

چکیده

background: despite several reports regarding the use of the occlutech figulla® flex septal occluder (offso) in adults, there are few reports on its use in children. we sought to study the result of the transcatheter closure of atrial septal defect (asd) using the offso in children ≤ 12 years. methods: w e enrolled 45 consecutive patients, ranging from 2.5 to 12 years of age, in two large pediatric cardiovascular centers. all the children underwent complete echocardiographic examination before the procedure. defect/device ratio and device/weight ratio were measured. the device diameter to the cardiac diameter ratio (dd/cd ratio) in anteroposterior projection after device release and the dd/cd index were calculated by dividing the dd/cd ratio by the body surface area. results: of the 45 enrolled patients, 25 (55%) were female. the range and mean ± standard deviation (sd) of age were2.5 to 12 years and 6.8 ± 2.5 years, respectively. the range and mean ± sd weight were 8.5 to 37.0 kg and 19.7 ± 7.2 kg, respectively. successful implantation was performed in all the patients. no major complications occurred in any of the subjects. we encountered one cobra head deformity in one patient. neither residual shunt nor conduction abnormality was observed in any of the cases. conclusion: t ranscatheter asd closure using the offso was effective in our pediatric patients. although this device needs relatively larger delivery sheaths, its use is safe while closing even large defects in children.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transcatheter Atrial Septal Defect Closure Using Occlutech Figulla Device: A Two-Center Experience

BACKGROUND Despite several reports regarding the use of the Occlutech Figulla® Flex septal occluder (OFFSO) in adults, there are few reports on its use in children. We sought to study the result of the transcatheter closure of atrial septal defect (ASD) using the OFFSO in children ≤12 years. METHODS We enrolled 45 consecutive patients, ranging from 2.5 to 12 years of age, in two large pediatr...

متن کامل

Percutaneous closure of a postoperative residual atrial septal defect with the Occlutech Figulla Occluder device.

A 67-year-old male patient with an eight-month history of operation for mitral valve repair and secundum atrial septal defect (ASD) presented with complaints of fatigue and shortness of breath. Transthoracic echocardiography showed a residual ASD resulting from separation of a pericardial patch. Qp/Qs rate was 3.2. The diameter of the residual defect measured by transesophageal echocardiography...

متن کامل

The efficiency and safety of percutaneous closure of secundum atrial septal defects with the Occlutech Figulla device: initial clinical experience.

OBJECTIVES We evaluated the efficiency and safety of the Occlutech Figulla device in percutaneous closure of secundum atrial septal defects (ASD). STUDY DESIGN The study included 28 patients (17 women, 11 men; mean age 43 years) who underwent percutaneous transcatheter closure using the Occlutech Figulla device for secundum ASDs causing a hemodynamically significant shunt. Defect size was est...

متن کامل

Transcatheter Closure of Secundum Atrial Septal Defect Using the Amplatzer Device: Single Center Experience in 140 Patients

In this paper we present our experience with the Amplatzer septal occluder device, employed in 140 patients for percutaneous closure of atrial secundum defect (ASD), from October 2002 to February 2006. The age of patients ranged between 5.3 and 70 years, median 21.9 years. Procedure time ranged between 25 and 240 minutes, median 60 minutes; fluoroscopy time ranged between 3.5 and 45 minutes, me...

متن کامل

[Transcatheter closure of atrial septal defect].

Isolated secundum atrial septal defect is one of the most common congenital heart defects. Surgical closure is the treatment of choice but is associated with a chest scar, some morbidity and a relatively long recovery and the use of cardiopulmonary bypass. Transcatheter closure of secundum atrial septal defect is therefore an attractive approach. 3 children, aged 5-10 years, underwent successfu...

متن کامل

The Occlutech Figulla PFO and ASD occluder: a new nitinol wire mesh device for closure of atrial septal defects.

BACKGROUND We investigated the safety, feasibility and efficacy of the Occlutech devices for patent foramen ovale (PFO) and atrial septal defect (ASD) closure in a prospective trial. METHODS The retention discs of the PFO device (23 x 25 mm, 27 x 30 mm) are connected by a 3 mm waist in the center with only one right atrial side central pin. The left atrial disc is produced either with a singl...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
the journal of tehran university heart center

جلد ۸، شماره ۴، صفحات ۱۹۷-۲۰۱

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023