cyclic fatigue resistance and fractographic analysis of race and protaper rotary niti instruments

نویسندگان

shahram azimi

parisa delvari

hamid cyrus hajarian

mohammad ali saghiri

چکیده

introduction: this study investigated the fatigue and fracture modes of race and protaper rotary instruments. materials & methods: fatigue resistance was evaluated by rotating the files 30 ° or 60 ° and with 2 or 5mm radius of curvature. race taper 06 size 25 and protaper f1 files (n=40) were used. the number of rotations to failure was analyzed by two-way anova and independent sample     t-test. fracture surfaces were examined under a scanning electron microscope. results: both files exhibited significantly more resistance to fracture when angle severity was reduced and increase in curvature radius (p=0.000). protaper demonstrated higher number of cycle of failure (p=0.0029) in one group (r=5mm, 60 º ). sem observation revealed fatigue mark/features in 17 specimens, shear characteristics in 37 samples, and tensile overload in 26 samples. conclusion: radius of curvature was the main factor in torsional and fatigue failures.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cyclic Fatigue Resistance and Fractographic Analysis of Race and Protaper Rotary NiTi Instruments

INTRODUCTION This study investigated the fatigue and fracture modes of RaCe and ProTaper rotary instruments. MATERIALS AND METHODS Fatigue resistance was evaluated by rotating the files 30° or 60° and with 2 or 5mm radius of curvature. RaCe taper 06 size 25 and ProTaper F1 files (n=40) were used. The number of rotations to failure was analyzed by two-way ANOVA and independent sample t-test. F...

متن کامل

Effect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments.

The resistance of ProTaper (Dentsply Maillefer, Ballaigues, Switzerland) nickel-titanium rotary instruments to cyclic fatigue was examined after their initial use in straight or curved canals in vivo. These instruments were rotated freely inside a steel phantom until separation. The number of rotations before failure and the lengths of the separated fragments were compared with data derived fro...

متن کامل

In-vitro Comparison of NiTi Tee, RaCe and Protaper Instruments in Dentinal Crack Formation

Background and Aim:Root canal preparation with rotary instruments may cause dentinal cracks leading to tooth fracture. The aim of this study was to compare three different rotary systems ProTaper, RaCe and Niti Tee on formation of dentinal cracks following root canal preparation. Materials and Methods: In this experimental study, 50 extracted mandibular first molars were selected. Teeth having ...

متن کامل

Cyclic fatigue resistance of rotary NiTi instruments after simulated clinical use in curved root canals.

The aim of this study was to assess cyclic fatigue resistance in rotary nickel-titanium instruments after simulated clinical use in curved root canals (40-degree, 5-mm radius curve). Thirty-six RaCe rotary NiTi files, size #5, taper 0.04, were divided into 3 groups: Groups A, B and C with one, three and five cycles of use, respectively. Time to failure was recorded with a stopwatch in seconds a...

متن کامل

Differences in cyclic fatigue resistance between ProTaper Next and ProTaper Universal instruments at different levels.

INTRODUCTION New designs and alloys have been developed to increase cyclic fatigue (CF) resistance of rotary files. The aim of this study was to compare CF resistance of ProTaper Universal (PTU; Dentsply Tulsa Dental, Tulsa, OK) and ProTaper Next (PTN, Dentsply Tulsa Dental) instruments at different points of curvature. METHODS A total of 420 files (240 PTU, S1, F1, F2, and F3 and 180 PTN, X1...

متن کامل

Cyclic fatigue of NiTi rotary instruments in a simulated apical abrupt curvature.

AIM To evaluate the cyclic fatigue resistance of five NiTi rotary systems in an abrupt apical curvature. METHODOLOGY Cyclic fatigue testing was performed in stainless-steel artificial canals with a 2-mm radius of curvature and an angle of curvature of 90 degrees constructed to the dimensions of the instruments tested. The middle of the simulated curvature was 2.5 mm from the tip of the instru...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
iranian endodontic journal

جلد ۶، شماره ۲، صفحات ۸۰-۸۵

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023