comparison between clinically used irregular fields shielded by cerrobend and standard lead blocks

نویسندگان

a r farajollahi

f bouzarjomehri

m kiani

چکیده

normal 0 false false false en-us x-none fa introduction: in radiation therapy centers across iran, protection of normal tissues is usually accomplished by either cerrobend or lead block shielding. in this study, the influence of these two shielding methods on central axis dose distribution of photon beam a cobalt unit was investigated in clinical conditions. materials and methods: all measurements were performed for 60 co γ-ray beams and the cerrobend blocks were fabricated by commercial cerrobend materials. standard lead block shields belonged to cobalt unit. data was collected through a calibrated ionization chamber, relative dosimetry systems and a tld dosimetery. results: results of the percent depth dose (pdd) measurements at depths of 0.5, 1, 5, 10, 15 and 20 cm for 23 different field sizes of patients with head and neck cancer showed no significant differences between lead and cerrobend shielding methods. measurement results of absolute dosimetry in depths of 1.5, 3, 5, 7, 10 and 12 cm also showed no significant differences between these two shielding methods. the same results were obtained by tld dosimetry on patient skin.   conclusion: use of melt shielding methods is a very easy and fast shield-making technique with no differences in pdd, absolute and skin dose between lead and cerrobend block shielding methods.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison between Clinically Used Irregular Fields Shielded by Cerrobend and Standard Lead Blocks

Introduction: In radiation therapy centers across Iran, protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. In this study, the influence of these two shielding methods on central axis dose distribution of photon beam a Cobalt unit was investigated in clinical conditions. Materials and Methods: All measurements were performed for 60Co γ-ray beams an...

متن کامل

Comparison between Clinically Used Irregular Fields Shielded by Cerrobend and Standard Lead Blocks

INTRODUCTION In radiation therapy centers across Iran, protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. In this study, the influence of these two shielding methods on central axis dose distribution of photon beam a Cobalt unit was investigated in clinical conditions. MATERIALS AND METHODS All measurements were performed for 60Co γ-ray beams and...

متن کامل

Effect of Material and Wall Thickness Buildup Caps on the Head ScatterFactor Measurements in Irregular Fields Shielded by Cerrobend

Introduction: The head scatter factor (Sc) is important to measurements radiation beam and beam modeling of treatment planning systems used for advanced radiation therapy techniques. This study aimed to investigate the design of a mini-phantom to measurement variations in collimator Sc in the presence of shielding blocks for shaping the beam using different field sizes. <str...

متن کامل

Evaluation of the physical characteristic of Cerrobend blocks used for radiation therapy

Background: In Radiation Therapy, the sparing of normal tissues can be performed using either multi-leaf collimators or Cerrobend blocks. The current work focuses on the physical characteristics of Cerrobend blocks including attenuation coefficient, effective penumbra width and isodose curves undulation in penumbral regions. Materials and Methods: All measurements were performed using ...

متن کامل

comparison of lead and cerrobend blocks for incident photon flux of 6 and 15 mv x-rays

background: during radiotherapy treatment, critical organs are shielded using lead and cerrobend blocks. the objective of this study is to compare the effects of lead and cerrobend shielding blocks on incident photon beam. methods: collimator scatter factors were measured for open square fields (3 x 3 cm to 40 x 40 cm) defined by collimator jaws and for fields blocked down to smaller asymmetric...

متن کامل

Comparison between small radiation therapy electron beams collimated by Cerrobend and tubular applicators

The purpose of this study was to compare the dosimetric properties of small field electron beams shaped by circular Cerrobend blocks and stainless steel tubular applicators. Percentage depth dose curves, beam profiles, and output factors of small-size circular fields from 2 to 5 cm diameter, obtained either by tubular applicators and Cerrobend blocks, were measured for 6, 10, and 15 MeV electro...