A transobturator adjustable system for male incontinence: 30-month follow-up of a multicenter study.

نویسندگان

  • Salomon Victor Romano
  • Wilhelm Huebner
  • Flavio Trigo Rocha
  • Fernando Pires Vaz
  • Valter Muller
  • Fabio Nakamura
چکیده

PURPOSE To report long-term results of the Argus T adjustable system for treatment of post-prostatectomy urinary incontinence (PPI). MATERIALS AND METHODS From October 2007 to August 2008, 37 patients with PPI were included in a prospective, single-arm, multicenter trial of treatment with the Argus T adjustable system (Promedon, Argentina). Preoperative evaluation included urine culture, urethrocystoscopy, urodynamic testing, 24-h pad weight test (PWT) and quality of life questionnaires. Patients were stratified according to baseline degree of incontinence (mild-moderate or severe). Postoperative evaluation included immediate PWT, quality of life questionnaires and daily use of pads at 1, 12 and 30 months. RESULTS AND CONCLUSIONS One patient was lost to follow-up. At the 30-month follow-up, 24/31 patients (77%) were dry, 3/31 (10%) improved and 4/31 (13%) were failures. In particular, in the mild-moderate group, 8/8 (100%) patients were dry. In the severe group, 20/28 patients (71%) were dry, 3/28 (11%) improved and 5/28 (18%) were failures. Median visual analogue scale (VAS) scores dropped from 9 (4-10) to 0.5 (0-10) and International Consultation on Incontinence Questionnaire Short Form scores from (ICIQ-SF) 19 (12-21) to 1 (0-10). Retrograde leak point pressure increased from 18 (5-29) to 35 (22-45) cm H2O after intraoperative adjustment. Complications included immediate postoperative infection in 2/36 patients (6%) and transient inguinal and/or perineal pain in 22/36 patients (61%). Argus T has a long-term high success rate (86% cure + improvement at the 30-month follow-up). Good outcomes were achieved even in severe incontinence cases and maintained for over 30 months.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Adjustable Transobturator Tape (Tot) For Post- Prostatectomy Incontınence

Introduction: Post-prostatectomy stress urinary incontinence (PPI) is a serious complication that affects patients’ quality of life. There is no pharmacological treatment approved for male stres urinary incontinence. The artificial urinary sphincter is the current ‘gold standard’ for the surgical treatment of incontinence after PPI. In this retrospective study, we aimed to evaluate the efficien...

متن کامل

Male incontinence and the transobturator approach: An analysis of current outcomes

OBJECTIVES To provide an analysis of the outcomes in patients who have a suburethral sling placed using the transobturator approach for the treatment of stress urinary incontinence after radical prostatectomy, because data in this specific area remain limited, and recent changes in male sling surgery might improve the efficacy in men with moderate or mild incontinence. METHODS We evaluated th...

متن کامل

Transobturator male sling TOMS for the treatment of stress post-prostatectomy incontinence, initial experience and results with one year's experience.

PURPOSE Post-prostatectomy incontinence remains a problem, even in minor or moderate degrees. In order to minimize surgical morbidity and costs, sling procedures have been proposed. The authors have developed a new transobturator male sling procedure and report their results after one-year experience. MATERIALS AND METHODS A prospective multicenter study was conducted in 50 patients with mino...

متن کامل

Comparison of TOA and TOT for Treating Female Stress Urinary Incontinence: Short-Term Outcomes

PURPOSE The transobturator adjustable tape (TOA) sling operation is a new procedure that allows for the adjustment of tension after surgical intervention, thus permitting correction of postoperative incontinence or obstruction. The aim of this study was to compare the efficacy and safety of TOA with that of the transobturator tape (TOT) procedure. MATERIALS AND METHODS Between 2008 and 2009, ...

متن کامل

First experience with the ATOMS® implant, a new treatment option for male urinary incontinence

INTRODUCTION Urinary incontinence (UI) is defined as any complaint of involuntary urine leakage. A description is provided of our experience with the ATOMS(®) (Adjustable Transobturator Male System. Agency for Medical Innovations. A.M.I.) adjustable implant in patients with mild to moderate UI. MATERIAL AND METHODS A retrospective study was made of the data referring to 13 patients treated wi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • International braz j urol : official journal of the Brazilian Society of Urology

دوره 40 6  شماره 

صفحات  -

تاریخ انتشار 2014