Glaucoma-Induced Optic Disc Morphometric Changes and Glaucoma Diagnostic Ability of Heidelberg Retina Tomograph II in Highly Myopic Eyes
نویسندگان
چکیده
PURPOSE This study was performed to first investigate the morphological differences in the optic nerve head between highly myopic non-glaucomatous controls and highly myopic glaucomatous eyes in comparison with the differences between emmetropic non-glaucomatous controls and emmetropic glaucomatous eyes using confocal scanning laser ophthalmoscopy. Further, the ability of the apparatus in glaucoma diagnosis in highly myopic eyes was compared with that in emmetropic eyes. METHODS Healthy subjects and age-matched patients with early-stage open-angle glaucoma were divided into two groups: emmetropic eyes (-1.0 to +1.0 diopters) and highly myopic eyes (-12.0 to -5.0 diopters).The participants were comprised of 65 emmetropic normal eyes, 59 emmetropic glaucomatous eyes, 62 highly myopic normal eyes, and 68 highly myopic glaucomatous eyes and eyes with pathologic myopia were carefully excluded. Confocal scanning laser tomographic parameters were compared among all subjects after adjustment for age and disc area. The ROC curves and sensitivity and specificity for glaucoma detection using several clinical methods were then compared between the emmetropic and highly myopic eyes. RESULTS Rim area, cup/disc area ratio, mean cup depth, and cup shape measure of glaucoma eyes are significantly different from those of normal eyes in both highly myopic eyes and emmetropic eyes. Methodological overestimation of retinal nerve fiber layer cross sectional area due to optic disc tilting was suggested in the highly myopic eyes. The diagnostic performance of glaucoma using several discriminant methods significantly deteriorated in the highly myopic eyes. CONCLUSIONS In the highly myopic glaucomatous eyes, confocal scanning laser tomographic parameters were significantly different from that of non-glaucomatous highly myopic eyes but diagnostic performance of glaucoma was deteriorated than that in emmetropic eyes. These findings demonstrate the utility and limitations of the apparatus in diagnosing glaucoma in highly myopic patients.
منابع مشابه
Influence of myopic disc shape on the diagnostic precision of the Heidelberg Retina Tomograph.
PURPOSE To investigate the diagnostic capability of a glaucoma diagnostic classification program for the Heidelberg Retina Tomograph (HRT) in eyes with myopic disc shapes. METHODS Sixty-six normal subjects (66 eyes) and 78 open-angle glaucoma patients (78 eyes) were enrolled. The eyes were divided into two groups; those eyes with myopic and those with nonmyopic disc shapes. The classification...
متن کاملCorrelation between morphology of optic disc determined by Heidelberg Retina Tomograph II and visual function in eyes with open-angle glaucoma
PURPOSE To investigate the correlation between the morphology of the optic disc and the visual function of eyes with open-angle glaucoma (OAG). PATIENTS AND METHODS One hundred and three eyes of 76 patients with OAG were studied. The baseline optic disc morphology was used to classify the eyes into four types: focal ischemic type (FI), myopic glaucomatous type (MY), senile sclerotic type (SS)...
متن کاملPeripapillary Intrachoroidal Cavitation and Optic Pit in a Case of High Myopia and Glaucoma
Purpose: To report a case of peripapillary intrachoroidal cavitation (choroidal schisis) and optic pit in a case of high myopia and open-angle glaucoma. Case report: A 30-year-old high myopic man was referred to our clinic for presumed peripapillary choroidal tumors. In funduscopy, there were two yellow-orange and slightly elevated peripapillary lesions in both eyes with indistinct margins. A s...
متن کاملEffect of Photorefractive Keratectomy on Optic Nerve Head Topography and Retinal Nerve Fiber Layer Thickness Measured by Heidelberg Retina Tomograph 3
PURPOSE To determine whether photorefractive keratectomy (PRK) has a significant effect on optic nerve head (ONH) parameters and peripapillary retinal nerve fiber layer (RNFL) thickness measured by the Heidelberg Retina Tomograph 3 (Heidelberg Engineering GmbH, Heidelberg, Germany) in eyes with low to moderate myopia. METHODS This prospective, interventional case series, includes 43 consecuti...
متن کاملEffect of disease severity and optic disc size on diagnostic accuracy of RTVue spectral domain optical coherence tomograph in glaucoma.
PURPOSE To evaluate the effect of disease severity and optic disc size on the diagnostic accuracies of optic nerve head (ONH), retinal nerve fiber layer (RNFL), and macular parameters with RTVue (Optovue, Fremont, CA) spectral domain optical coherence tomography (SDOCT) in glaucoma. METHODS 110 eyes of 62 normal subjects and 193 eyes of 136 glaucoma patients from the Diagnostic Innovations in...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 9 شماره
صفحات -
تاریخ انتشار 2014