Aseptic lymphocyte-dominated vasculitis-associated lesion: a clinicopathologic review of an underrecognized cause of prosthetic failure.
نویسندگان
چکیده
It is estimated that 35% of total hip arthroplasties (THAs) involve a second-generation metal-on-metal (MOM) prosthesis. A novel complication has appeared in a subset of patients with MOM THAs that is described as an aseptic, lymphocyte-dominated vasculitis-associated lesion (ALVAL). The clinical features of ALVAL are nonspecific, but patients complain of pain and may develop "pseudotumors." It is hypothesized that metal ions are released from the prosthesis and form haptens with native proteins that elicit a type IV hypersensitivity response in the local soft tissues. Histopathologic descriptions of ALVAL are similar to those of failed arthroplasty in general, with the addition of a dense perivascular inflammatory infiltrate that is the hallmark of ALVAL. We report 3 cases of ALVAL with clinical, radiographic, and histologic findings. Accurate assessment is crucial because an intraoperative diagnosis of chronic inflammation suggestive of ALVAL will necessitate a replacement of the prosthetic component surfaces.
منابع مشابه
Periprosthetic osteolysis due to metastatic renal cell carcinoma: a case report
Failure of total hip arthroplasty through septic or aseptic loosening, periprosthetic fracture, or recurrent dislocation is well recognized and understood. We present an unusual cause of failure of total hip replacement which occurred on a 79 year old gentleman: that of prosthetic loosening secondary to malignant infiltration around components. Our aim is to highlight the fact that malignant in...
متن کاملPrevalence of adverse reactions to metal debris following metal-on-metal THA.
The purpose of this study was to determine the prevalence of adverse reactions to metal debris (ARMD) following large-diameter metal-on-metal total hip arthroplasty. The authors examined the potential for using magnetic resonance imaging to screen for pseudotumors in 108 hips 2 years postoperatively. Serum cobalt and chromium concentrations were measured in 80 hips that underwent unilateral tot...
متن کاملParticle Disease: A Current Review of the Biological Mechanisms in Periprosthetic Osteolysis After Hip Arthroplasty
BACKGROUND Inflammatory responses to wear debris cause osteolysis that leads to aseptic prosthesis loosening and hip arthroplasty failure. Although osteolysis is usually associated with aseptic loosening, it is rarely seen around stable implants. Aseptic implant loosening is a simple radiologic phenomenon, but a complex immunological process. Particulate debris produced by implants most commonl...
متن کاملIs oral vitamin D a risk factor for prosthetic loosening?
High counts of wear particles and activated macrophages are strong predictors of prosthetic loosening in total joint arthroplasty patients. The tissue response, dominated by macrophages, with production of inflammatory mediators and matrix-degrading enzymes, triggers a self-accelerating cycle of osteolysis, ultimately resulting in failure of the arthroplasty, high treatment cost and poor patien...
متن کاملThe correlation of wear with histological features after failed hip resurfacing arthroplasty.
BACKGROUND Tissue necrosis and a macrophage and perivascular lymphocytic infiltrate are commonly seen in periprosthetic tissues around metal-on-metal hip resurfacing implants, including pseudotumors associated with these implants. The purpose of the present study was to correlate pathological changes in periprosthetic tissues with clinical findings and the amount of implant-derived metal wear. ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- American journal of clinical pathology
دوره 134 6 شماره
صفحات -
تاریخ انتشار 2010