Automated particle-counting immunoassay for digoxin.

نویسندگان

  • D Collet-Cassart
  • C G Magnusson
  • C L Cambiaso
  • M Lesne
  • P L Masson
چکیده

We describe a non-radioisotopic automated immunoassay for serum digoxin in the therapeutic range at 50 samples per hour by use of the particle-counting immunoassay technique and a special reagent, rheumatoid factor. Digoxin is determined by evaluating its inhibitory activity on the agglutination of digoxin-coated latex particles. The agglutinating system consists of the anti-digoxin IgG mixed with rheumatoid factor. This anti-IgG autoantibody reacts with IgG when they are aggregated or bound to a surface. Interference by serum proteins was prevented by pepsin digestion at 37 degrees C. Correlation with radioimmunoassay was r = 0.94 for 109 samples with concentrations between 0 and 5.7 micrograms/L. Analytical recovery overall averaged 102%. Within-assay precision (CV) ranged from 12.5% at 0.96 micrograms/L to 4.4% at 2.97 micrograms/L; between-assay precision varied from 8.4% at 0.75 micrograms/L to 5.8% at 2.92 micrograms/L.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Abbott TDx, Dade Stratus, and Du Pont aca automated digoxin immunoassays compared with a reference radioimmunoassay method.

Automated digoxin immunoassays with the Abbott TDx, Dade Stratus, and Du Pont aca were evaluated against an RIA similar to the Centers for Disease Control Candidate Reference Method. All three automated methods correlated poorly with the reference method. Particularly disconcerting were the falsely increased values (up to 0.6 micrograms/L) obtained by the Stratus for blank specimens. We discuss...

متن کامل

Immunoassay of digoxin by differential centrifugation.

We describe an immunoassay of digoxin that exploits the differential sedimentation rate between two types of latex particles in a microcentrifugal analyzer. Conventional polystyrene latex particles (relative density = 1.05), sensitized with antibody, are first suspended with the sample and then mixed with digoxin-sensitized latex particles (relative density = 1.5). In the absence of digoxin in ...

متن کامل

Endogenous digoxin-like immunoreactive factors eliminated from serum samples by hydrophobic silica-gel extraction and enzyme immunoassay.

Elimination of endogenous digoxin-like immunoreactive factors (DLIF) that interfere with accurate measurement of digoxin requires use of a highly specific anti-digoxin antibody, or that DLIF be separated from digoxin before immunoassay. Several commercial digoxin-assay kits include a step for separating serum proteins and other substances from digoxin before immunoassay. We tested six different...

متن کامل

Intercomparison of seven radioimmunoassay kits and a fluorescence polarization immunoassay kit for digoxin.

Samples from 33 patients being treated with digoxin and three concentrations of control material were analyzed for this drug by use of seven radioimmunoassay (RIA) kits and an automated fluorescence polarization immunoassay (FPIA) system. CVs ranged from 2.4 to 6.4% for the FPIA system and from 4.6 to 7.4% for two RIA methods. In the analysis of between-method variability of RIA kits, CVs range...

متن کامل

Automated particle-counting immunoassay of C-reactive protein and its application to serum, cord serum, and cerebrospinal fluid samples.

The assay for C-reactive protein has been fully automated as a particle-counting immunoassay. For cerebrospinal fluid, cord serum, and adults' serum its range of sensitivity extends from 1 microgram/L to 300 mg/L, with a minimal sample dilution of twofold and a maximal dilution of 50-fold being required. This range is so broad because free antibodies are added to the reaction medium. However, w...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 27 7  شماره 

صفحات  -

تاریخ انتشار 1981