Spatially constrained incoherent motion method improves diffusion-weighted MRI signal decay analysis in the liver and spleen.
نویسندگان
چکیده
PURPOSE To evaluate the effect of the spatially constrained incoherent motion (SCIM) method on improving the precision and robustness of fast and slow diffusion parameter estimates from diffusion-weighted MRI in liver and spleen in comparison to the independent voxel-wise intravoxel incoherent motion (IVIM) model. METHODS We collected diffusion-weighted MRI (DW-MRI) data of 29 subjects (5 healthy subjects and 24 patients with Crohn's disease in the ileum). We evaluated parameters estimates' robustness against different combinations of b-values (i.e., 4 b-values and 7 b-values) by comparing the variance of the estimates obtained with the SCIM and the independent voxel-wise IVIM model. We also evaluated the improvement in the precision of parameter estimates by comparing the coefficient of variation (CV) of the SCIM parameter estimates to that of the IVIM. RESULTS The SCIM method was more robust compared to IVIM (up to 70% in liver and spleen) for different combinations of b-values. Also, the CV values of the parameter estimations using the SCIM method were significantly lower compared to repeated acquisition and signal averaging estimated using IVIM, especially for the fast diffusion parameter in liver (CVIV IM = 46.61 ± 11.22, CVSCIM = 16.85 ± 2.160, p < 0.001) and spleen (CVIV IM = 95.15 ± 19.82, CVSCIM = 52.55 ± 1.91, p < 0.001). CONCLUSIONS The SCIM method characterizes fast and slow diffusion more precisely compared to the independent voxel-wise IVIM model fitting in the liver and spleen.
منابع مشابه
Spatially Constrained Incoherent Motion (SCIM) Model Improves Quantitative Diffusion-Weighted MRI Analysis of Crohn's Disease Patients
Quantitative analysis of fast and slow diffusion from abdominal Diffusion-weighted MRI has the potential to provide important new insights into physiological and microstructural properties of the body. However, the commonly used, independent voxel-wise fitting of the signal decay model leads to imprecise parameter estimates, which has hampered their practical usage. In this work we evaluated th...
متن کاملSpatially Constrained Probability Distribution Model of Incoherent Motion (spim) Improves Longitudinal Reproducibility of Quantitative Diffusion Weighted Mri
Purpose: Diffusion-weighted MRI (DW-MRI) of the body is a non-invasive imaging technique that enables characterization of tissue microenvironments through measurement of variations in the mobility of water molecules due to cellularity, cell membrane integrity, and compartment they are located. Quantitative DW-MRI has a well-established role in the detection, characterization and follow-up of a ...
متن کاملA Spatially Constrained Probability Distribution Model of Incoherent Motion (spim) in Quantitative Diffusion Weighted Mri
Purpose: Quantitative diffusion-weighted MRI (DW-MRI) has a well-established role in the detection and characterization of a variety of abdominal abnormalities including liver fibrosis, tumors and active inflammation. DW-MRI enables characterization of the tissue microstructure through measurement of the mobility of water molecules and the variations in their mobility due to changes in cell den...
متن کاملReliable estimation of incoherent motion parametric maps from diffusion-weighted MRI using fusion bootstrap moves
Diffusion-weighted MRI has the potential to provide important new insights into physiological and microstructural properties of the body. The Intra-Voxel Incoherent Motion (IVIM) model relates the observed DW-MRI signal decay to parameters that reflect blood flow in the capillaries (D*), capillaries volume fraction (f), and diffusivity (D). However, the commonly used, independent voxel-wise fit...
متن کاملIntravoxel incoherent motion in body diffusion-weighted MRI: reality and challenges.
OBJECTIVE Diffusion-weighted MRI is increasingly applied in the body. It has been recognized for some time, on the basis of scientific experiments and studies in the brain, that the calculation of apparent diffusion coefficient by simple monoexponential relationship between MRI signal and b value does not fully account for tissue behavior. However, appreciation of this fact in body diffusion MR...