Differential recovery of the electroretinogram, visually evoked cortical potential, and electrically evoked cortical potential following vitrectomy: implications for acute testing of an implanted retinal prosthesis.

نویسندگان

  • Sandra R Montezuma
  • Joseph F Rizzo
  • Ofer R Ziv
چکیده

To determine the extent to which electrophysiologic tests of the afferent visual pathway are affected by vitrectomy, the procedure was performed in 15 eyes of 11 adult Dutch-belted rabbits. An electroretinogram (ERG), visually evoked cortical potential (VECP), and electrically evoked cortical potential (EECP) were obtained preoperatively and sequentially after surgery. For electrical stimulations, biphasic impulses were delivered to the retina. Post-vitrectomy declines of 49, 25, and 41% from the median baseline amplitudes and increases of 13, 18, and 17% from the median baseline latency values were found for ERG, VECP, and EECP, respectively. At 90 min, 13 to 30% of eyes still had an amplitude more than 10% below baseline on at least one of the three tests, whereas 10 to 47% of eyes had an abnormal latency more than 10% above baseline on at least one of the three tests. Amplitudes were more likely than latencies to return to near baseline, but for eyes that remained subnormal, the decline was greater for amplitudes than latencies. Significant alterations in retinal function, manifested by declines in amplitudes and increases in latencies of the ERG, VECP, and EECP, persist in a large proportion of eyes up to 90 min post-vitrectomy.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the Wave Amplitude of Visually Evoked Potential in Amblyopic Eyes between Patients with Esotropia and Anisometropia and a Normal Group

Background: We compared the wave amplitude of visually evoked potential (VEP) between patients with esotropic and anisometropic amblyopic eyes and a normal group.Methods: The wave amplitude of VEP was documented in 2 groups of persons with amblyopia (15 with esotropia and 28 with anisometropia) and 1 group of individuals with normal visual acuity (n, 15). The amplitude of P100 was recorded mono...

متن کامل

Visual evoked potential in rabbits’ eyes with subretinal implantation by vitrectomy of Okayama University-type retinal prosthesis (OURePTM)

Okayama University-type retinal prosthesis (OURePTM) is a photoelectric dye-coupled polyethylene film which generates electric potential in response to light and stimulates nearby neurons. This study aims to test surgical feasibility for subretinal film implantation and to examine functional durability of films in subretinal space. Dye-coupled films were implanted subretinally by vitrectomy in ...

متن کامل

Evaluation of the Slope of Amplitude Growth Function Changes of the Electrically Evoked Action Potential in Three Months after Receiving the Device in Children with Cochlear Implant

Objective: In neural response telemetry, intracochlear electrodes stimulate the auditory nerve and record the neural responses. The electrical stimulation is sent to the auditory nerve by an electrode and the resulted response, called electrically evoked compound action potential, is recorded by an adjacent electrode. The most important clinical applications of this test are evaluation and moni...

متن کامل

Relationship between Electrically Evoked Compound Action Potential Thresholds and Auditory, Language, and Speech Progress after Cochlear Implant Surgery

Introduction: Electrically evoked compound action potential (ECAP) is an objective auditory response that can be used in the programing of cochlear implants. The aims of this study were to monitor ECAP thresholds and auditory, language and speech progress for 6 months after cochlear implant surgery and to evaluate any relationship between them.   Materials and Methods: Ten children with a m...

متن کامل

Fabrication of Pillar Shaped Electrode Arrays for Artificial Retinal Implants

Polyimide has been widely applied to neural prosthetic devices, such as the retinal implants, due to its well-known biocompatibility and ability to be micropatterned. However, planar films of polyimide that are typically employed show a limited ability in reducing the distance between electrodes and targeting cell layers, which limits site resolution for effective multi-channel stimulation. In ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of rehabilitation research and development

دوره 41 2  شماره 

صفحات  -

تاریخ انتشار 2004