Present state and issues in IORT Physics
نویسنده
چکیده
Literature was reviewed to assess the physical aspects governing the present and emerging technologies used in intraoperative radiation therapy (IORT). Three major technologies were identified: treatment with electrons, treatment with external generators of kV X-rays and electronic brachytherapy. Although also used in IORT, literature on brachytherapy with radioactive sources is not systematically reviewed since an extensive own body of specialized literature and reviews exists in this field. A comparison with radioactive sources is made in the use of balloon catheters for partial breast irradiation where these are applied in almost an identical applicator technique as used with kV X-ray sources. The physical constraints of adaption of the dose distribution to the extended target in breast IORT are compared. Concerning further physical issues, the literature on radiation protection, commissioning, calibration, quality assurance (QA) and in-vivo dosimetry of the three technologies was reviewed. Several issues were found in the calibration and the use of dosimetry detectors and phantoms for low energy X-rays which require further investigation. The uncertainties in the different steps of dose determination were estimated, leading to an estimated total uncertainty of around 10-15% for IORT procedures. The dose inhomogeneity caused by the prescription of electrons at 90% and by the steep dose gradient of kV X-rays causes additional deviations from prescription dose which must be considered in the assessment of dose response in IORT.
منابع مشابه
Evaluating the performance of a 3D PLA buildup bolus in breast intraoperative radiotherapy.
Introduction: The received dose to the surface of tumor bed is one of the main issues in beast intraopeative radiotherapy (IORT). Totally, the administered dose to the surface is usually lower than that of radiotherapy reference point (distal end of tumor bed). Employing a buildup bolus can effectively increase the surface dose and therefore, eliminate such created cold points ...
متن کاملIntraoperative radiation therapy using mobile electron linear accelerators: report of AAPM Radiation Therapy Committee Task Group No. 72.
Intraoperative radiation therapy (IORT) has been customarily performed either in a shielded operating suite located in the operating room (OR) or in a shielded treatment room located within the Department of Radiation Oncology. In both cases, this cancer treatment modality uses stationary linear accelerators. With the development of new technology, mobile linear accelerators have recently becom...
متن کاملBeam characteristics and leakage assessment of an in-house intra-operative electron applicator system
Background: Intraoperative radiation therapy (IORT) technique is used to treat a surgically exposed tumor or tumor bed in the treatment of locally advanced intraabdominal, retroperitoneal, pelvic, thoracic, breast and soft tissue tumors. One way to perform IORT is to use an existing linac and modify the treatment head using an applicator system. In this study the objectives were to design, and ...
متن کاملIntraoperative radiotherapy using a mobile electron LINAC: A retroperitoneal sarcoma case
The advent of mobile LINACs for use in intraoperative radiation therapy (IORT) promises to make IORT more accessible than before and easier to deliver to patients undergoing surgery. Although mobile IORT systems have been available since 1999, few treatment centers currently use them. Here, we present the case of a typical patient undergoing IORT for retroperitoneal sarcoma to show how easy the...
متن کاملCommissioning the First Mobile Dedicated Accelerator for Intraoperative Electron Radiotherapy in Iran
Introduction: Intraoperative radiotherapy is a radiotherapy technique in which a high single fraction of radiation dose is delivered to the patient after surgery and Concurrent with anesthesia time. The most frequent method for IORT implementation is Intraoperative electron radiotherapy (IOERT), in which, some dedicated and high dose per pulse electron accelerators are employe...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 12 شماره
صفحات -
تاریخ انتشار 2017