COBAS AMPLICOR: fully automated RNA and DNA amplification and detection system for routine diagnostic PCR.
نویسندگان
چکیده
The COBAS AMPLICOR system automates amplification and detection of target nucleic acids, making diagnostic PCR routine for a variety of infectious diseases. The system contains a single thermal cycler with two independently regulated heating/cooling blocks, an incubator, a magnetic particle washer, a pipettor, and a photometer. Amplified products are captured on oligonucleotide-coated paramagnetic microparticles and detected with use of an avidin-horseradish peroxidase (HRP) conjugate. Concentrated solutions of amplicon or HRP were pipetted without detectable carryover. Amplified DNA was detected with an intraassay CV of < 4.5%; the combined intraassay CV for amplification and detection was < 15%. No cross-reactivity was observed when three different target nucleic acids were amplified in a single reaction and detected with three target-specific capture probes. The initial COBAS AMPLICOR menu includes qualitative tests for diagnosing infections with Chlamydia trachomatis, Neisseria gonorrhoeae, Mycobacterium tuberculosis, and hepatitis C virus. All tests include an optional Internal Control to provide assurance that specimens are successfully amplified and detected.
منابع مشابه
Evaluation of the automated COBAS AMPLICOR hepatitis C virus PCR system.
To evaluate the reliability and feasibility of the automated Roche COBAS AMPLICOR PCR system for routine detection of hepatitis C virus (HCV) RNA, a total of 405 serum samples previously tested by an in-house nested PCR and manual Roche AMPLICOR microwell plate HCV test were examined. Complete concordance was found between the results with the HCV COBAS AMPLICOR system and the previously determ...
متن کاملEvaluation of quantification of HIV-1 RNA viral load in plasma and dried blood spots by use of the semiautomated Cobas Amplicor assay and the fully automated Cobas Ampliprep/TaqMan assay, version 2.0, in Kisumu, Kenya.
In Kenya, HIV-1 viral load monitoring is commonly performed with the Cobas Amplicor using plasma specimens. Interest is growing in transitioning to real-time PCR (RT-PCR), such as the Cobas Ampliprep/Cobas TaqMan (CAP/CTM), using dried blood spots (DBS). Before implementation, direct evaluation of the two assays using DBS field specimens is required. This study compares the sensitivity, specifi...
متن کاملFully automated detection of hepatitis C virus RNA in serum and whole-blood samples.
In this study, we established a fully automated molecular assay for qualitative detection of hepatitis C virus (HCV) in serum and whole-blood samples and compared it with conventional molecular assays, including manual HCV RNA extraction protocols. Whole-blood samples were collected from patients with and without chronic HCV infection in EDTA tubes and nucleic acid stabilization tubes (NASTs). ...
متن کاملMulticenter evaluation of the COBAS AMPLICOR HCV assay, an integrated PCR system for rapid detection of hepatitis C virus RNA in the diagnostic laboratory.
The benefits shown by the recent introduction of PCR for the in vitro diagnosis of hepatitis C virus (HCV) infection has prompted the development of standardized, ready-to-use assays that can be implemented in routine clinical laboratories. We have evaluated the clinical performance of COBAS AMPLICOR HCV (COBAS), the first instrument system that allows the automation of HCV RNA amplification an...
متن کاملAutomated specific capture of hepatitis C virus RNA with probes and paramagnetic particle separation.
We developed and evaluated a prototype automated specimen preparation instrument for the specific capture of hepatitis C virus (HCV) RNA with probes and magnetic bead-fluid separation. HCV RNA was isolated from serum by lysis of virus particles with a chaotropic agent, followed by hybridization of the RNA with biotinylated probes and capture of the hybridized RNA with streptavidin-coated parama...