Eddy current-induced artifact reduction in spiral trajectory balanced-SSFP for transition-band SSFP fMRI
نویسندگان
چکیده
J. Lee, J. M. Pauly Electrical Engineering, Stanford University, Stanford, CA, United States Introduction Balanced SSFP is known to be very sensitive to any magnetization disturbance since all gradients are fully refocused in every TR and the phase accrual between RF pulses solely comes from offresonance precession. One of the major sources for this magnetization disturbance is eddy currents that are induced from changing spiral interleaves or phase encodes. Different spiral interleaves or phase encodes induce different amounts of eddy currents that can cause additional spin dephasing (due to B0 component of the eddy currents) in balanced SSFP. As a result, serious image artifacts can be observed based on spiral interleaves or phase encodes ordering (Fig. 1a) [1]. Bieri et al. proposed an eddy currents compensation technique for RF phase-cycled Cartesian balanced-SSFP sequence [1]. Here we proposed an eddy-currents compensation method for non-RF phase-cycled 3D stack-of-spiral balanced-SSFP sequence. The formation of eddy current-induced artifacts is illustrated and a new interleaves ordering method is presented. Major applications for this method are transition-band SSFP fMRI [2] and real-time respiration compensation technique [3]. Theory and Methods In conventional balanced SSFP where RF phase cycling is utilized, the sequential phase encoding order in Cartesian or rotational interleaf order in the interleaved spiral trajectory (Fig. 3a,c) mitigates the eddy currentinduced artifacts. The additional spin dephasing introduced from the eddy currents of the previous phase encode or spiral interleaf is cancelled out by the similar amount of dephasing from the next phase encode or spiral interleaf. This is because the adjacent phase encodes or spiral interleaves have a very small amount of magnetization change and the RF phase cycling shifts the direction of the induced eddy currents in every TR (Fig. 2a). Therefore no significant artifact can be observed. However, in transition-band SSFP fMRI, where smallflip-angle balanced SSFP with no RF phase cycling is used, the eddy currents induced from the sequential ordering cause serious artifacts in the image (Fig. 1a). In the sequential ordering scheme, since the RF no longer flips the dephasing direction of the eddy currentinduced magnetization, the amount of spin dephasing is piled up until the phase encode or spiral interleaf magnetization goes to the opposite direction (Fig. 2b,c). This additional dephasing causes imaging artifacts as shown in Fig 1a. Hence, a new phase or interleaves ordering scheme is required to remove the artifacts. One simple and effective method is named as the ‘alternating spiral interleaves (phase encodes) ordering’. This method pairs the opposite polarity spiral interleaves (phase encodes) and sequences them next to each other (Fig. 3b,d). These two opposite interleaves (phase encodes) create opposing eddy currents that will cancel themselves out in every other TR. As a result, the additional spin dephasing effect does not persist anymore (Fig 2d). In the real-time respiration compensation method [3], this alternating order is critical in measuring a reliable respiration-induced resonance offset. Since the respiration-induced resonance offset is measured from the phase of FID, the eddy current-induced spin dephasing degrades the measurement accuracy. Results The resulting image from the alternating ordering scheme shows reduced imaging artifacts (Fig. 1b). The alternating ordering also provides significantly reduced eddy current-induced spin dephasing in the FID phase measurement (Fig. 4). Discussion and Conclusion A new interleaves (phase encodes) ordering method for non-phase-cycled balanced SSFP is presented and tested for a 3D stack-of-spiral sequence. With this method, the eddy current induced artifacts were clearly reduced and the respiration was measured with less noise. The proposed method can also be used to cover concentric trajectories because after an alternating pair the next phase encode can start anyplace as long as it is paired with its alternating phase encode one. Hence it can provide a similar phase encodes ordering flexibility suggested by [1]. References [1] Bieri O. MRM 54:129-137 (2005) [2] Miller KL. MRM 50:675-683 (2003) [3] Lee J. ISMRM Miami. p.100 (2005)
منابع مشابه
Analysis of the BOLD Signal Characteristics in balanced SSFP fMRI: a Monte-Carlo Simulation
Introduction Recently, balanced SSFP (Steady State Free Precession) has been proposed as an alternative method to measure the BOLD contrast [1-3]. Two different contrast mechanisms have been explored experimentally: (1) transition-band SSFP fMRI (BOSS fMRI) that detects the BOLD induced frequency shift near the transition part of the SSFP profile [1,2], and (2) pass-band SSFP fMRI that is belie...
متن کاملDouble Average Parallel Imaging for Optimized Eddy Current Compensation and Steady State Storage in balanced SSFP Imaging
M. Markl, J. Leupold, J. Hennig Diagnostic Radiology, Medical Physics, University Hospital Freiburg, Freiburg, Germany Introduction: k-space segmented time-resolved (CINE) balanced SSFP imaging is widely used in cardiac MR-imaging [1]. Drawbacks include sensitivity to the disruption of the steady state and limitations regarding non-linear k-space reordering which is problematic for SSFP imaging...
متن کاملReal-time B0 field drift compensation in balanced SSFP: Stabilization of the activation band in transition-band SSFP fMRI
J. Lee, J. M. Santos, J. M. Pauly Electrical Engineering, Stanford University, Stanford, CA, United States Introduction In fMRI experiments, the presence of slow B0 resonance frequency drift is often observed. Several factors, including instrumental instability [1], physiological shifts and subject motion-related-noise are suspected for the causes of this drift. Henry et al. [2] reported that t...
متن کاملFMRI using balanced steady-state free precession (SSFP)
Steady-state free precession (SSFP) is a highly-efficient MRI pulse sequence that has been a fairly recent arrival in the functional MRI realm. Several methods for using balanced SSFP to detect the BOLD signal have been proposed to date and will be discussed in this review. After a brief introduction to the general properties of SSFP, this review describes the quite different approaches of tran...
متن کاملEnhanced spectral shaping in steady-state free precession imaging.
Balanced steady-state free precession (SSFP) is hindered by the inherent off-resonance sensitivity and unwanted bright fat signal. Multiple-acquisition SSFP combination methods, where multiple datasets with different fixed RF phase increments are acquired, have been used for shaping the SSFP spectrum to solve both problems. We present a new combination method (weighted-combination SSFP or WC-SS...