BACKGROUND This work presents a generalized technique to estimate pulmonary ventilation-to-volume (v/V) distributions using the multiple-breath nitrogen washout, in which both tidal volume (V T ) and the end-expiratory lung volume (EELV) are allowed to vary during the maneuver. In addition, the volume of the series dead space (v d ), unlike the classical model, is considered a common series unit connected to a set of parallel alveolar units. METHODS The numerical solution for simulated data, either error-free or with the N2 measurement contaminated with the addition of Gaussian random noise of 3 or 5 % standard deviation was tested under several conditions in a computational model constituted by 50 alveolar units with unimodal and bimodal distributions of v/V. Non-negative least squares regression with Tikhonov regularization was employed for parameter retrieval. The solution was obtained with either unconstrained or constrained (V T , EELV and v d ) conditions. The Tikhonov gain was fixed or estimated and a weighting matrix (WM) was considered. The quality of estimation was evaluated by the sum of the squared errors (SSE) (between reference and recovered distributions) and by the deviations of the first three moments calculated for both distributions. Additionally, a shape classification method was tested to identify the solution as unimodal or bimodal, by counting the number of shape agreements after 1000 repetitions. RESULTS The accuracy of the results showed a high dependence on the noise amplitude. The best algorithm for SSE and moments included the constrained and the WM solvers, whereas shape agreement improved without WM, resulting in 97.2 % for unimodal and 90.0 % for bimodal distributions in the highest noise condition. CONCLUSIONS In conclusion this generalized method was able to identify v/V distributions from a lung model with a common series dead space even with variable V T . Although limitations remain in presence of experimental noise, appropriate combination of processing steps were also found to reduce estimation errors.
منابع مشابه
Generalized estimation of the ventilatory distribution from the multiple-breath washout: a bench evaluation study
BACKGROUND The multiple-breath washout (MBW) is able to provide information about the distribution of ventilation-to-volume (v/V) ratios in the lungs. However, the classical, all-parallel model may return skewed results due to the mixing effect of a common dead space. The aim of this work is to examine whether a novel mathematical model and algorithm is able to estimate v/V of a physical model,...
متن کاملEvaluation of the Impact of Alveolar Nitrogen Excretion on Indices Derived from Multiple Breath Nitrogen Washout
BACKGROUND A large body of evidence has now accumulated describing the advantages of multiple breath washout tests over conventional spirometry in cystic fibrosis (CF). Although the majority of studies have used exogenous sulphur hexafluoride (SF6) as the tracer gas this has also led to an increased interest in nitrogen washout tests, despite the differences between these methods. The impact of...
متن کاملMoment ratio analysis of multiple breath nitrogen washout in infants with lung disease.
Measurement of lung volumes at end expiratory level and assessment of ventilation inhomogeneity is important for respiratory management in infants with lung disease. This study compared multiple breath nitrogen washout was compared with body plethysmography to measure functional residual capacity in infants and assessed ventilation inhomogeneity using mean dilution numbers and alveolar based ga...
متن کاملDefining the appropriate waiting time between multiple-breath nitrogen washout measurements.
Static lung volume measurements allow objective assessment of total lung capacity (TLC), functional residual capacity (FRC) and residual volume, and assist in the diagnosis and management of lung disease [1, 2]. The use of gas dilution techniques, such as multiple-breath nitrogen washout (MBNW), require minimal patient cooperation and allow FRC to be determined in those individuals unable to co...
متن کاملMultiple-breath nitrogen washout techniques: including measurements with patients on ventilators.
Definition ....................................................................... Methods for determining FRC ................................... Choice of method FRC measured by plethysmography (FRCpleth) ......... FRC measured by helium dilution (FRCHe) ............. FRC measured by nitrogen washout (FRCN2) ........... Standards for lung volume measurement by gas dilution/washout techniques ...