Organ doses and subjective image quality of indirect digital panoramic radiography.
نویسندگان
چکیده
OBJECTIVES To determine if slight variations in exposure will affect diagnostic image quality and absorbed radiation doses for digital and analogue panoramic radiography. METHODS Thermoluminescent dosimeters were placed in the thyroid gland, eyes, submandibular glands, parotid glands and skin of two human cadaver heads. Three different exposure settings were used: 70 kV, 120 mAs; 77 kV, 75 mAs; and 81 kV, 60 mAs. Subjective image quality was assessed using a phantom head. Storage phosphor (SP) images were printed on film and both analogue and SP images were assessed for their subjective image quality on a five-point rating scale. The results were statistically analysed using logistic regression analysis and chi(2) tests. RESULTS Highest organ doses were measured for the submandibular glands, followed by the parotid glands. Salivary gland doses tended to be higher at lower kV settings. Image quality was not statistically different for the different exposure settings. Imaging technique did not seem to influence diagnostic image quality, except for the periapical status of upper premolars where SP was better. The main reason for any differences appeared to be interobserver variation. CONCLUSIONS Analogue and SP panoramic radiography performed equally well for subjective diagnostic image quality. No significant differences could be found at the exposure settings used in this study. Radiation doses were highest for the salivary glands, especially at lower kV settings.
منابع مشابه
Quality comparison of direct digital panoramic radiography and computed radiography panoramic
BACKGROUND AND AIM: Digital panoramic X-ray images can be captured using photostimulable phosphors or solid-state detectors (i.e. charge-coupled devices and Flat-Panels). The first category is defined as computed radiography (CR) or semi-direct radiography. The second technology that uses solid-state detectors is known as direct digital radiography (DDR). Both of these technologies have their o...
متن کاملImage quality enhancement in digital panoramic radiograph
One of the most common positioning errors in panoramic radiography is palatoglossal air space above the apices of the root of maxillary teeth. It causes a radiolucency obscuring the apices of maxillary teeth. In the case of this positioning error, the imaging should be repeated. This causes the patient be exposed to radiation again. To avoid the repetition of exposing harmful X-rays to the pati...
متن کاملThe Practice of Chest Radiography Using Different Digital Imaging Systems: Dose and Image Quality
Introduction: The study was undertaken to evaluate the practice of chest radiography using different digital imaging systems and its influence on dose and image quality. Materials and Methods: The study was carried out in two hospitals from March 2016 to June 2016. Sixty ambulatory patients aged 21 to 60 years who were able to cooperate without difficulty and weighed between 60 to 80 kg were ...
متن کاملEvaluation of Image Magnification in Digital Panoramic Radiography
Introdouction: The aim of the present study was to investigate image magnification in digital panoramic radiography and assess the effect of anatomical sites on the magnification in both males and females. Materials and methods: In this retrospective study, digital panoramic radiographs of 48 patients with 202 implants were investigated. Implant diameter and length were measured on pan...
متن کاملEvaluation of the Influence of Exposure Index on Image Quality and Radiation Dose
Introduction: The introduction of digital radiography has led to a significant problem in terms of dose creep. To address this problem, manufacturers have established a set of exposure indicators (EI) as a feedback mechanism to safeguard against overexposure. The EI is the measure of incident exposure to the detector that is directly proportional to the signal-to-noise...