Chronic intravascular hemolysis after aortic valve replacement. Long-term study comparing different types of ball-valve prostheses.

نویسندگان

  • E Eyster
  • J Rothchild
  • O Mychajliw
چکیده

To assess the severity of hemolysis and to determine whether severity of hemolysis was related to type of prosthesis, hematocrit readings (Hct), schistocyte counts (S), reticulocyte counts (R), and determinations of haptoglobin (Hp), lactic acid dehydrogenase (LDH), bilirubin, urinary hemoglobin, and urinary hemosiderin (Hs) were performed on 54 patients who had three different types of aortic ball-valve prostheses. Hemolysis, as determined by absence of haptoglobins or hemosiderinuria or R > 3.5% in the absence of bleeding, was detected in 39 of the 54 patients (72%). S, R, LDH, and bilirubin were higher in patients without haptoglobins than in those with haptoglobins (P < 0.005). Hematocrits of less than 32% were found in 13 of the 34 patients (38%) with Starr-Edwards valves compared to none of the 11 with Magovern valves (P <0.001) and one of the nine with Cutter (P < 0.05) valves. The combination of S > 10 per 1,000 cells, R> 5%, and LDH> 500 mU was observed only with Starr-Edwards valves (16 of the 34 patients) and was not seen in patients with Magovem or Cutter valves (P < 0.005). All patients with S > 10, R > 5%, and LDH > 500 mU had hemosiderinuria, and five had hemoglobinuria. Hemoglobinuria was always indicative of severe hemolysis. Hemosiderinuria was rarely detected less than 3 months after operation but was found in 32 of 47 patients so tested (68%) more than 3 months after operation. Hemosiderinuria was found in 25 of 30 patients with StarrEdwards valves compared to two of the 11 with Magovern valves (P < 0.001). Hemosiderinuria was also detected in five of six patients with Cutter valves tested, but in three of these the test was only faintly positive. It is concluded that hemolysis is more severe in the presence of Starr-Edwards valves than Magovern or Cutter valves. Based on these observations, the following criteria for rapid clinical evaluation of cardiac hemolysis are suggested: Mild: Hs or absence of Hp but S < 10, R < 5, LDH < 500; moderate: Hs or absence of Hp but S > 10, R > 5, LDH > 500; severe: all of above plus hemoglobinuria.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Late postoperative anatomic observations after insertion of Hufnagel caged-ball prostheses in descending thoracic aorta.

Clinical and necropsy observations are described in two patients who died 11 and 13 years, respectively, after implantation of Hufnagel prostheses (the first ever used successfully in humans to treat cardiac valve disease) in descending thoracic aorta. Despite the long implantation periods, there was no evidence of prosthetic degeneration or thrombosis or intravascular hemolysis. Although appro...

متن کامل

Intravascular haemolysis after valve replacement: comparative study between Starr-Edwards (ball valve) and Björk-Shiley (disc valve) prosthesis.

Seventy-four patients with single prosthetic valves (Björk-Shiley or Starr-Edwards) in the mitral or aortic position and 18 controls with rheumatic valvar heart disease were investigated for evidence of intravascular haemolysis. Serum lactate dehydrogenase (LDH) was used as the most sensitive indicator of haemolysis. Raised concentrations were found in a third of 39 patients with Björk-Shiley p...

متن کامل

Chronic hemolysis following fascia lata and Starr-Edwards aortic valve replacement.

Sixty-two patients were investigated for chronic intravascular hemolysis from 3 months to 6iS years following aortic valve replacement; 27 patients had Starr-Edwards prostheses and 35 fascia lata grafts. The findings are suggestive of a significantly higher incidence and degree of hemolysis in patients with Starr-Edwards prostheses compared to those with fascia lata grafts. In both groups regur...

متن کامل

Traumatic hemolysis with hemoglobinuria due to ball variance.

P OSTIMPLANTATION ALTERATION of the physicochemical properties of the silastic poppet ( ball variance ) has been reported in some patients with aortic ball valve prostheses.12 Clinical recognition of this situation is difficult, and the diagnosis is usually made at autopsy. This report concerns a patient with ball variance manifested primarily by traumatic intravascular hemolysis and hemoglobin...

متن کامل

Intravascular haemolysis and urinary iron losses after replacement of heart valves by a prosthesis.

1. The haemolytic complications of Starr-Edwards ball and cage prostheses have been studied in ninety-five patients. 2. Intravascular haemolysis has been found to be common and particularly marked after an aortic valve replacement. 3. A simple and reliable method for estimating urinary iron is described, and excessive urinary iron loss was found to be common especially in aortic valve replaceme...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Circulation

دوره 44 4  شماره 

صفحات  -

تاریخ انتشار 1971