A clinical comparison of Xe-127 and Xe-133 for ventilation studies.
نویسندگان
چکیده
Xenon-133 and xenon-127 were compared by performing ventilation studies with both radionuclides in 19 patients with a variety of lung diseases. Assessment of the counting rate over the chest, relative to the radioactivity in the lungs, permitted the evaluation of each isotope in terms of usable photons detected by a scintillation camera with a large field of view and appropriate collimation. A greater photon yield was obtained with Xe-127. Markedly improved resolution was shown by measurement of a line phantom, but was not apparent on subjective appraisal of scintiphotos except in the washout phase. Xenon-127 appears to be preferable to Xe-133 because of the higher counting rates, lower patient radiation dose, and longer shelf life. In addition, a prior perfusion study using a Tc-99m radiopharmaceutical does not affect the quality of a Xe-127 ventilation study. The use of Xe-127 therefore permits the selection for ventilation studies of only those patients with suspected pulmonary embolism, and eliminates unnecessary radiation exposure. A further improvement in image quality obtained with Xe-127 should be possible with certain modifications of the scintillation camera that would permit use of the 375-keV photopeak along with the 172- and 203-keV gamma energies. Charcoal traps designed for Xe-133 will require additional shielding and longer storage time when used for Xe-127. Xenon-127, however, might be used again after appropriate processing.
منابع مشابه
jnm / DIAGNOSTIC NUCLEAR MEDICINE
Xenon-133 and xenon-127 were compared by performing ventilation studies with both radionuclides in 19 patients with a variety of lung diseases. Assessment of the counting rate over the chest, relative to the radioactivity in the lungs, permitted the evaluation of each isotope in terms of usable photons detected by a scintillation camera with a large field of view and ap propriate collimili ion....
متن کاملRegional ventilation studies with Kr-81m and Xe-133: a comparative analysis.
Krypton-81m and Xe-133 ventilation scintigrams, and particulate per fusion scintigrams with Tc-99m, were obtained on the same day in 41 pa tients. Of 14 studies demonstrating pulmonary embolism, Kr-81m corn plernented the perfusion study better than Xe-133 in seven and was judged essential for that diagnosis in two. Xenon-133 was better in none. On the other hand, in 29 studies with obstructive...
متن کاملComparison of Xe-133 washout and single-breath imaging for the detection of ventilation abnormalities.
The Xe-133 ventilation studies of 115 patients were analyzed to determine the relative abilities of the single-breath and washout phases to detect regional ventilation abnormalities. All Xe-133 images were obtained in the posterior projection before 6-view perfusion studies with Tc-99m human albumin microspheres. There were 275 regions with matching V-P abnormalities in the patients. The washou...
متن کاملComparison of Quantitative Ventilation-Perfusion Lung Scan Results Using Different Xenon-133 Ventilation Images
Objective: This paper evaluates the influence of the specific Xe ventilation images used to calculate differential function in quantitative ventilation-perfusion lung scans. Methods: Whole-lung differentials were determined on 43 Xe ventilationfl mTc-macroaggregated albumin perfusion studies. Ventilation calculations were performed for the firstbreath, up to three equilibrium and the composite ...
متن کامل133Xe contamination found in internal bacteria filter of xenon ventilation system.
OBJECTIVE We report on (133)Xe contamination found in the reusable internal bacteria filter of our xenon ventilation system. METHODS Internal bacteria filters (n = 6) were evaluated after approximately 1 mo of normal use. The ventilation system was evacuated twice to eliminate (133)Xe in the system before removal of the filter. Upon removal, the filter was monitored using a survey meter with ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of nuclear medicine : official publication, Society of Nuclear Medicine
دوره 18 7 شماره
صفحات -
تاریخ انتشار 1977