Micromotion of dental implants may interfere with the process of osseointegration. Using three different types of virtual biomechanical models, varying contact types between implant and bone were simulated, and implant deformation, bone deformation, and stress at the implant-bone interface were recorded under an axial load of 200 N, which reflects a common biting force. Without friction between implant and bone, a symmetric loading situation of the bone with maximum loading and displacement at the apex of the implant was recorded. The addition of threads led to a decrease in loading and displacement at the apical part, but loading and displacement were also observed at the vertical walls of the implants. Introducing friction between implant and bone decreased global displacement. In a force fit situation, load transfer predominantly occurred in the cervical area of the implant. For freshly inserted implants, micromotion was constant along the vertical walls of the implant, whereas, for osseointegrated implants, the distribution of micromotion depended on the location. In the cervical aspect some minor micromotion in the range of 0.75 μm could be found, while at the most apical part almost no relative displacement between implant and bone occurred.
منابع مشابه
Novel Method for Testing Cementless Tibial Components Using Micromotion and Impedance Vibration Analysis
INTRODUCTION: The initial mechanical stability of cementless implants is critical to achieve successful osseointegration through minimization of deleterious micromotion. Controversy exists regarding the most reliable and appropriate methodology for in vitro mechanical testing of implants used to guide design and performance of cementless total knee replacement. Mechanical testing methods vary w...
متن کاملDesign improvement and dynamic finite element analysis of novel ITI dental implant under dynamic chewing loads.
The main aim of this article was to introduce the application of a uniform design for experimental methods to drop the micromotion of a novel ITI dental implant model under the dynamic loads. Combining the characteristics of the traditional ITI and Nano-Tite implants, a new implant with concave holes has been constructed. Compared to the traditional ITI dental implant model, the micromotion of ...
متن کاملEffects of different implant-abutment connections on micromotion and stress distribution: prediction of microgap formation.
OBJECTIVES The aim of this study was to analyse micromotion and stress distribution at the connections of implants and four types of abutments: internal hexagonal, internal octagonal, internal conical and trilobe. METHODS A three dimensional (3D) model of the left posterior mandible was reconstructed from medical datasets. Four dental implant systems were designed and analysed independently i...
متن کاملEvaluating the impact of length and thread pitch on the stress distribution in dental implants and surrounding bone using finite element method
longevity of osseointegrated implants are intensely influenced by biomechanical factors. Control of these factors prevents mechanical complications, which include fracture of screws, components, or materials veneering the framework. In this study, the impact of length and threads pitch of dental implants on the stress distribution and maximum Von Mises stress in implant-abutment complex and ja...
متن کاملAn Overview of the Mechanical Integrity of Dental Implants
With the growing use of dental implants, the incidence of implants' failures grows. Late treatment complications, after reaching full osseointegration and functionality, include mechanical failures, such as fracture of the implant and its components. Those complications are deemed severe in dentistry, albeit being usually considered as rare, and therefore seldom addressed in the clinical litera...