Computational comparison of aortic root stresses in presence of stentless and stented aortic valve bio-prostheses.
نویسندگان
چکیده
We provide a computational comparison of the performance of stentless and stented aortic prostheses, in terms of aortic root displacements and internal stresses. To this aim, we consider three real patients; for each of them, we draw the two prostheses configurations, which are characterized by different mechanical properties and we also consider the native configuration. For each of these scenarios, we solve the fluid-structure interaction problem arising between blood and aortic root, through Finite Elements. In particular, the Arbitrary Lagrangian-Eulerian formulation is used for the numerical solution of the fluid-dynamic equations and a hyperelastic material model is adopted to predict the mechanical response of the aortic wall and the two prostheses. The computational results are analyzed in terms of aortic flow, internal wall stresses and aortic wall/prosthesis displacements; a quantitative comparison of the mechanical behavior of the three scenarios is reported. The numerical results highlight a good agreement between stentless and native displacements and internal wall stresses, whereas higher/non-physiological stresses are found for the stented case.
منابع مشابه
Stentless aortic valves. Current aspects
The design of stentless valve prostheses is intended to achieve a more physiological flow pattern and superior hemodynamics in comparison to stented valves. First - generation stentless bioprosthesis were the Prima valve, the Freestyle valve and the Toronto stentless porcine valve. The second generation of stentless valves, as the Super stentless aortic porcine valve, need only one suture line....
متن کاملFull-root aortic valve replacement with stentless xenograft achieves superior regression of left ventricular hypertrophy compared to pericardial stented aortic valves
BACKGROUND Full-root aortic valve replacement with stentless xenografts has potentially superior hemodynamic performance compared to stented valves. However, a number of cardiac surgeons are reluctant to transform a classical stented aortic valve replacement into a technically more demanding full-root stentless aortic valve replacement. Here we describe our technique of full-root stentless aort...
متن کاملStented bioprostheses in aortic position
INTRODUCTION Biological stented prostheses are currently the main type of prosthetic valve used for aortic valve replacement. The ratio of bioprotheses to mechanical prostheses has switched in the last 15 years; the percentage of biological prostheses implanted has risen from 30 % to 85 %. Moreover the total number of implanted stented bioprostheses remained stable over the last years despite ...
متن کاملStentless Aortic Valve/Root Replacement
In-vitro, stentless valves have a larger effective orifice area than their stented counterparts. This is best exemplified by the fact that a 25-mm Mosaic valve, which is stented, is manufactured using a 23-mm Freestyle stentless valve. So, in order to end-up with similar effective orifice areas when implanting a stentless valve or a stented valve in the same size left ventricular outflow tract,...
متن کاملHaemodynamics and left ventricular mass regression: a comparison of the stentless, stented and mechanical aortic valve replacement.
OBJECTIVE Our objective was to compare the degree of change in hemodynamics and left ventricular mass (LVM) regression after aortic valve replacement (AVR) with stentless, stented and mechanical valves. METHODS Patients greater than 59 years of age had AVR for aortic stenosis with the stentless xenograft (Cryolife-O'Brien, CLOB), stented xenograft (Carpentier-Edwards, C-E) or mechanical valve...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Computer methods in biomechanics and biomedical engineering
دوره 20 2 شماره
صفحات -
تاریخ انتشار 2017