An Algorithmic Readout Approach for Thermal Conductivity Based CO2 Sensors
نویسندگان
چکیده
This thesis presents a new approach to reading out thermal-conductivitybased gas sensors. This method is intended for the readout of a CMOScompatible resistive thermal-conductivity transducer for indoor CO2 sensing applications, without requiring precision off-chip components. Instead of accurately regulating the power dissipated in the transducer and measuring its temperature, the temperature and power dissipation are both measured using an algorithmic approach. A high-resolution ADC digitizes the voltage drop across the transducer and the current through it, measured using an on-chip reference resistor. Moreover, by digitizing several base-emitter voltages of an on-chip bipolar transistor, a precision bandgap voltage reference is constructed in the digital domain, and accurate information about the ambient temperature is obtained, which is used to temperature compensate the voltage reference and the reference resistor. Thus, all necessary ingredients are obtained to calculate the power dissipation and temperature of the transducer, from which the thermal conductivity of the surrounding air, and hence CO2 concentration, can be obtained. A prototype integrated circuit implementing this readout approach has been realized in 0.16μm CMOS. It has been tested in a climate chamber in combination with a platinum resistor mimicking the transducer. The digitally-constructed voltage reference has a temperature coefficient of 9ppm/◦C, while ambient temperature is sensed with accuracy of ±0.2◦C, with a resolution of 0.15◦C. The resistance readings have an inaccuracy ranging between -1mΩ to 4mΩ on a nominal resistance of about 100Ω (-10ppm 40ppm) with a resolution of around 2mΩ in the temperature range from 10◦C to 40◦C; for the power measurements, the circuit achieved an accuracy between -0.03% and 0.06%, with an 800nW of resolution (in the same temperature range) which is one ii Abstract order of magnitude better than results presented in previous work. Although no CO2 measurements have been performed, an estimated thermalresistance accuracy of around 2862ppm with a resolution of 155.64[K/W] should be possible, which would enable detection of the CO2 levels in the air with an accuracy of around 0.72% and a resolution of 7705ppm.
منابع مشابه
Optimal SQUID based non-destructive test for detecting sub-surface defects with the help of advanced SQUID superconducting sensors and an experimental approach for optimal production method of these sensors from the YBCO superconductor materials
The conventional eddy current method for non-destructive inspection of welding joints has limitations that can examine defects to a certain depth below the surface of the sample and is not suitable for determining deep defects. This limitation can be overcome using the SQUID superconducting sensors. The nonstoichiometric composition of YBCO due to its superconducting temperature and desired cri...
متن کاملOptimal SQUID based non-destructive test for detecting sub-surface defects with the help of advanced SQUID superconducting sensors and an experimental approach for optimal production method of these sensors from the YBCO superconductor materials
The conventional eddy current method for non-destructive inspection of welding joints has limitations that can examine defects to a certain depth below the surface of the sample and is not suitable for determining deep defects. This limitation can be overcome using the SQUID superconducting sensors. The nonstoichiometric composition of YBCO due to its superconducting temperature and desired cri...
متن کاملHigh accuracy thermal conductivity measurement of aqueous cryoprotective agents and semi-rigid biological tissues using a microfabricated thermal sensor.
An improved thermal-needle approach for accurate and fast measurement of thermal conductivity of aqueous and soft biomaterials was developed using microfabricated thermal conductivity sensors. This microscopic measuring device was comprehensively characterized at temperatures from 0 °C to 40 °C. Despite the previous belief, system calibration constant was observed to be highly temperature-depen...
متن کاملSingle Walled Carbon Nanotube Effects on Mixed Convection heat Transfer in an Enclosure: a LBM Approach
The effects of Single Walled Carbon Nanotube (SWCNT) on mixed convection in a cavity are investigated numerically. The problem is studied for different Richardson numbers (0.1-10), volume fractions of nanotubes (0-1%), and aspect ratio of the cavity (0.5-2.5) when the Grashof number is equal to 103. The volume fraction of added nanotubes to Water as base fluid are lowers than 1% to make dilute ...
متن کاملTemporal Noise Analysis of Charge-Domain Sampling Readout Circuits for CMOS Image Sensors †
This paper presents a temporal noise analysis of charge-domain sampling readout circuits for Complementary Metal-Oxide Semiconductor (CMOS) image sensors. In order to address the trade-off between the low input-referred noise and high dynamic range, a Gm-cell-based pixel together with a charge-domain correlated-double sampling (CDS) technique has been proposed to provide a way to efficiently em...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2014