Sequence optimization to reduce velocity offsets in cardiovascular magnetic resonance volume flow quantification - A multi-vendor study
نویسندگان
چکیده
PURPOSE Eddy current induced velocity offsets are of concern for accuracy in cardiovascular magnetic resonance (CMR) volume flow quantification. However, currently known theoretical aspects of eddy current behavior have not led to effective guidelines for the optimization of flow quantification sequences. This study is aimed at identifying correlations between protocol parameters and the resulting velocity error in clinical CMR flow measurements in a multi-vendor study. METHODS Nine 1.5T scanners of three different types/vendors were studied. Measurements were performed on a large stationary phantom. Starting from a clinical breath-hold flow protocol, several protocol parameters were varied. Acquisitions were made in three clinically relevant orientations. Additionally, a time delay between the bipolar gradient and read-out, asymmetric versus symmetric velocity encoding, and gradient amplitude and slew rate were studied in adapted sequences as exploratory measurements beyond the protocol. Image analysis determined the worst-case offset for a typical great-vessel flow measurement. RESULTS The results showed a great variation in offset behavior among scanners (standard deviation among samples of 0.3, 0.4, and 0.9 cm/s for the three different scanner types), even for small changes in the protocol. Considering the absolute values, none of the tested protocol settings consistently reduced the velocity offsets below the critical level of 0.6 cm/s neither for all three orientations nor for all three scanner types. Using multilevel linear model analysis, oblique aortic and pulmonary slices showed systematic higher offsets than the transverse aortic slices (oblique aortic 0.6 cm/s, and pulmonary 1.8 cm/s higher than transverse aortic). The exploratory measurements beyond the protocol yielded some new leads for further sequence development towards reduction of velocity offsets; however those protocols were not always compatible with the time-constraints of breath-hold imaging and flow-related artefacts. CONCLUSIONS This study showed that with current systems there was no generic protocol which resulted into acceptable flow offset values. Protocol optimization would have to be performed on a per scanner and per protocol basis. Proper optimization might make accurate (transverse) aortic flow quantification possible for most scanners. Pulmonary flow quantification would still need further (offline) correction.
منابع مشابه
Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
AIMS Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error. ...
متن کاملBlood Flow Simulation in an Aorta with a mild coarctation Using Magnetic Resonance Angiography and Finite Volume Method
Coarctation of the aorta is one of the five main congenital cardiovascular failures, accounting for 6–8 percent of these failures. This research aimed to simulate the blood flow of a seventeen-year-old male teen with a mild coarctation at one-third of his aorta's descending branch. The simulation was performed by extracting the domain and the input pulsatile velocity signal as the boundary cond...
متن کاملOptimal bipolar gradient design to reduce velocity offsets
Introduction: Phase-contrast velocity quantification is a valuable tool in cardiac magnetic resonance imaging. However, eddy-currents cause additional signal phases, which result in offsets in the velocity measurements. Even a relative small offset (>0.6 cm/s) can result in significant errors in clinical parameters [1]. Gatehouse et al. [2] showed that for a typical breath-hold flow protocol no...
متن کاملPart Ii Velocity Offsets: Characterization in a Multi-vendor Study
Aims: Cardiovascular magnetic resonance (CMR) allows non-invasive phase contrast measurements of flow through planes transecting large vessels. However, some clinically valuable applications are highly sensitive to errors caused by small offsets of measured velocities if these are not adequately corrected, for example by the use of static tissue or static phantom correction of the offset error....
متن کاملBaseline correction of phase contrast images improves quantification of blood flow in the great vessels.
PURPOSE Phase-contrast Cardiovascular Magnetic Resonance Imaging (CMR) generally requires the analysis of stationary tissue adjacent to a blood vessel to serve as a baseline reference for zero velocity. However, for the heart and great vessels, there is often no stationary tissue immediately adjacent to the vessel. Consequently, uncorrected velocity offsets may introduce substantial errors in f...