A compact and stable eddy covariance set-up for methane measurements using off-axis integrated cavity output spectroscopy
نویسندگان
چکیده
A Fast Methane Analyzer (FMA) is assessed for its applicability in a closed path eddy covariance field set-up in a peat meadow. The FMA uses off-axis integrated cavity output spectroscopy combined with a highly specific narrow band laser for the detection of CH4 and strongly reflective mirrors to obtain a laser path length of 2–20×103 m. Statistical testing and a calibration experiment showed high precision (7.8×10−3 ppb) and accuracy (<0.30%) of the instrument, while no drift was observed. The instrument response time was determined to be 0.10 s. In the field set-up, the FMA is attached to a scroll pump and combined with a 3-axis ultrasonic anemometer and an open path infrared gas analyzer for measurements of carbon dioxide and water vapour. The power-spectra and co-spectra of the instruments were satisfactory for 10 Hz sampling rates. Due to erroneous measurements, spikes and periods of low turbulence the data series consisted for 26% of gaps. Observed CH4 fluxes consisted mainly of emission, showed a diurnal cycle, but were rather variable over. The average CH4 emission was 29.7 nmol m−2 s−1, while the typical maximum CH4 emission was approximately 80.0 nmol m−2 s−1 and the typical minimum flux was approximately 0.0 nmol m−2 s−1. The correspondence of the measurements with flux chamber measurements in the footprint was good and the observed CH4 emission rates were comparable with eddy covariance CH4 measurements in other peat areas. Additionally, three measurement techniques with lower sampling frequencies were simulated, which might give the possibility to measure CH4 fluxes without an external pump and save energy. Disjunct eddy covariance appeared to be the most reliable substitute for 10 Hz eddy covariance, while relaxed eddy accumulation gave reliable estimates of the fluxes over periods in the order of days or weeks. Correspondence to: D. M. D. Hendriks ([email protected])
منابع مشابه
Off-axis integrated cavity output spectroscopy with a mid-infrared interband cascade laser for real-time breath ethane measurements.
Cavity-enhanced tunable diode laser absorption spectroscopy is an attractive method for measuring small concentrations of gaseous species. Ethane is a breath biomarker of lipid peroxidation initiated by reactive oxygen species. A noninvasive means of quickly quantifying oxidative stress status has the potential for broad clinical application. We present a simple, compact system using off-axis i...
متن کاملSpatial and temporal dynamics in eddy covariance observations of methane fluxes at a tundra site in Northeastern Siberia1
In the past two decades, the eddy covariance technique has been used for an increasing number of methane flux studies at an ecosystem scale. Previously, most of these studies used a closed path setup with a tunable diode-laser spectrometer (TDL). Although this method worked well, the TDL has to be calibrated regularly and cooled with liquid nitrogen or a cryogenic system, which limits its use i...
متن کاملMeasurements of output factors using different ionization chambers and build up caps
Background: The aim of this work was to study the output factors for Linac using different ionization chambers and build up caps. Materials and Methods: Output factors were measured for open square fields (3 × 3 cm to 40 × 40 cm) defined by collimator jaws for 6 and 15 MV photon beams from a Varian Clinac 2100C accelerator were measured. The measurements in air were performed using Com...
متن کاملOptical parametric oscillator based off-axis integrated cavity output spectroscopy for rapid chemical sensing.
An optical parametric oscillator (OPO), pumped by a fiber-amplified diode laser, is combined with off-axis integrated cavity output spectroscopy (OA-ICOS). The cw OPO (power 1.2 W, tunability 3-4 μm, 5 cm(-1) mode-hop-free tuning) has a tuning speed of 100 THz/s, which is ideal for rapid and sensitive trace gas detection. Combined with OA-ICOS, a detection limit of 50 parts per trillion by volu...
متن کاملReal-time, subsecond, multicomponent breath analysis by Optical Parametric Oscillator based Off-Axis Integrated Cavity Output Spectroscopy.
Breath analysis is an attractive field of research, due to its high potential for non-invasive medical diagnostics. Among others, laser-based absorption spectroscopy is an excellent method for the detection of gases in exhaled breath, because it can combine a high sensitivity with a good selectivity, and a high temporal resolution. Here, we use a fast-scanning continuous wave, singly-resonant O...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2007