Comparison of pulse contour analysis by Pulsioflex and Vigileo to measure and track changes of cardiac output in critically ill patients.

نویسندگان

  • X Monnet
  • S Vaquer
  • N Anguel
  • M Jozwiak
  • F Cipriani
  • C Richard
  • J-L Teboul
چکیده

BACKGROUND We compared the new Pulsioflex and the Vigileo devices to measure cardiac index (CI) in critically ill patients. Both devices measure CI by pulse-contour analysis. The Pulsioflex device also allows an auto-calibration (not based on thermodilution). METHODS Patients were included if we administered fluids (20 patients), reduced (20 patients), or increased (20 patients) the dose of norepinephrine. Before and after interventions, we measured CI provided by the Vigileo (CIVig) and Pulsioflex (CIPfx) devices before and after its auto-calibration. CI measured by transpulmonary thermodilution (CIthermo) was used as the reference. RESULTS Considering absolute values of CI (n=120), the percentage error was 59% for CIVig vs CIthermo and 40% for CIthermo vs CIPfx. Auto-calibrating CIPfx after interventions did not improve the percentage error between CIPfx and CIthermo (39%). Considering the fluid-induced changes in CI, the coefficient of correlation with changes in CIthermo was 0.50 for CIVig, and 0.73 for CIPfx (P=0.27). It was not significantly improved if CIPfx was auto-calibrated (r=0.64). Considering the norepinephrine-induced changes in CI, the coefficient of correlation with changes in CIthermo was 0.41 for CIVig. It tended to be better for CIPfx (r=0.71, P=0.07). It was not significantly improved by auto-calibration (r=0.53). CONCLUSIONS The Pulsioflex did not reliably estimate the absolute values of CI. For tracking fluid-induced changes in CI, the Pulsioflex was reliable, and also the Vigileo. For tracking norepinephrine-induced changes in CI, it was also reliable and tended to be better than the Vigileo. Auto-calibration allowed by the system did not improve its reliability.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis

Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...

متن کامل

Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis.

BACKGROUND Recently, continuous monitoring of cardiac output (CO) based on pulse contour analysis (Vigileo) has been introduced into practice. In this clinical study, we evaluated the accuracy of this system by comparing it with the transpulmonary thermodilution technique (TPID) in septic patients. METHODS We studied 24 mechanically ventilated patients with septic shock (16 male, 8 female, ag...

متن کامل

Bioreactance is not reliable for estimating cardiac output and the effects of passive leg raising in critically ill patients.

BACKGROUND Bioreactance estimates cardiac output in a non-invasive way. We evaluated the ability of a bioreactance device (NICOM®) to estimate cardiac index (CI) and to track relative changes induced by volume expansion. METHODS In 48 critically ill patients, we measured CI estimated by the NICOM® device (CINicom) and by transpulmonary thermodilution (CItd, PiCCO2™ device) before and after a ...

متن کامل

Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in critically ill patients.

BACKGROUND The ability of the third-generation FloTrac/Vigileo software to track changes in cardiac index (CI) induced by volume expansion and norepinephrine in critically ill patients is unknown. METHODS In subjects with circulatory failure, we administered volume expansion (20 subjects) and increased (20 subjects) or decreased (20 subjects) the dose of norepinephrine. We measured arterial p...

متن کامل

Comparison between Flotrac-Vigileo and Bioreactance, a totally noninvasive method for cardiac output monitoring

INTRODUCTION This study was designed to compare the clinical acceptability of two cardiac output (CO) monitoring systems: a pulse wave contour-based system (FloTrac-Vigileo) and a bioreactance-based system (NICOM), using continuous thermodilution (PAC-CCO) as a reference method. METHODS Consecutive patients, requiring PAC-CCO monitoring following cardiac surgery, were also monitored by the tw...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • British journal of anaesthesia

دوره 114 2  شماره 

صفحات  -

تاریخ انتشار 2015