Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements.
نویسندگان
چکیده
It has been speculated that material loss, either as corrosion or wear, at the head-stem taper junction is implicated in the high revision rates reported for metal-on-metal total hip replacements. We measured the volume of material loss from the taper and bearing surfaces of retrieved devices, and investigated the associations with blood metal ion levels and the diagnosis of a cystic or solid pseudotumor. The median volumes of material lost from the female and male taper surfaces were 2.0 and 0.29 mm(3) , respectively, while the median volumes of wear from the cup and head bearing surfaces were 1.94 and 3.44 mm(3) , respectively. Material loss from the female taper was similar to that from the acetabular bearing surface (p = 0.55), but significantly less than that from the femoral bearing surface (p < 0.001). Material loss from the male taper was less than that from both bearing surfaces (p < 0.001). Multivariable analysis demonstrated no significant correlations between the volume of material lost from the taper surfaces and either blood cobalt or chromium ions, or the presence of pseudotumor. While a substantial volume of material is lost at the taper junction, the clinical significance of this debris remains unclear.
منابع مشابه
Assessing the material loss of the modular taper interface in retrieved metal on metal hip replacements
Measuring the amount of material loss in the case of revised hip replacements is considered to be a prerequisite of understanding and assessing the true in vivo performance of the implant. This paper outlines a method developed by the authors for quantifying taper material loss as well as more general taper interface parameters. Previous studies have mostly relied on visual inspection to assess...
متن کاملCorrosion on the acetabular liner taper from retrieved modular metal-on-metal total hip replacements.
Eight retrieved metal-on-metal total hip replacements displayed corrosion damage along the cobalt-chromium alloy liner taper junction with the Ti alloy acetabular shell. Scanning electron microscopy indicated the primary mechanism of corrosion to be grain boundary and associated crevice corrosion, which was likely accelerated through mechanical micromotion and galvanic corrosion resulting from ...
متن کاملTaper junction failure in large-diameter metal-on-metal bearings
OBJECTIVES An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer. METHODS Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular...
متن کاملClinically insignificant trunnionosis in large-diameter metal-on-polyethylene total hip arthroplasty
OBJECTIVES Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of materi...
متن کاملDamage Patterns at the Head-Stem Taper Junction Helps Understand the Mechanisms of Material Loss.
BACKGROUND Material loss at the taper junction of metal-on-metal total hip arthroplasties has been implicated in their early failure. The mechanisms of material loss are not fully understood; analysis of the patterns of damage at the taper can help us better understand why material loss occurs at this junction. METHODS We mapped the patterns of material loss in a series of 155 metal-on-metal ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of orthopaedic research : official publication of the Orthopaedic Research Society
دوره 31 11 شماره