The feasibility of the noninvasive assessment of blood 'coagulability' (the tendency to coagulate) has been tested by using a novel device, the Thrombo-Monitor. It monitors, by using the principles of near infra-red (NIR) dynamic light scattering, the tendency of blood to create clots. The Thrombo-Monitor observes the very initial changes of blood viscosity, which occurs due to the temporarily induced stasis of capillary blood of the finger. One hundred and fifteen patients aged >65 years (matched by age and sex) participated in the study. Patients were initially divided into four groups based on the patient's medical therapy. The study groups were: warfarin, enoxaparin, aspirin and/or clopidogrel, and a control group. The medications were given according to the patient's comorbidities (eg, atrial fibrillation [AF], status post pulmonary embolism [S/p PE], status post cerebrovascular accident [S/p CVA]). The Thrombo-Monitor Index (TMI) is a noninvasive index, derived on the basis of laboratory test results of international normalized ratio (INR) and prothrombin time (PT) values. For the group of patients who were treated only with warfarin, TMI was adjusted by using the jackknife statistical approach to create maximum correlation and linearity with INR and PT values that ranged from 1.1 to 5.0. For all warfarin patients (N = 35) the TMI was found to have a good correlation with INR and PT values (R(2) = 0.64, P < 0.00001); mean TMI = 1.86 (SD = 0.91); mean INR and PT = 2.3 (SD = 0.91). The calibration curve thus generated was used to calculate the TMI for all other groups: aspirin group, mean TMI = 1.3 (SD = 0.14, N = 23), corresponding approximately to INR and PT values of 1.036; enoxaparin group (N = 24), mean TMI = 1.34 (SD = 0.304), corresponding to mean INR and PT values of 1.07 (SD = 0.3); control group, INR and PT ≥ 1 (N = 32), mean TMI = 1.24 (SD = 0.32). R(2) of all control and warfarin patients (N = 67) was 0.55 (P < 0.00001). In summary, the newly introduced TMI index is significantly correlated with INR and PT values.
منابع مشابه
Developing and Validating a New Wireless Wearable Device for Balance Measurement in Sport and Clinical Setting
One of the new clinical techniques to assess the lower body parameters is the wearable ultrasonic sensors. The device which can measure the static and dynamic balance abilities in sport and clinical setting by the traveled signals of ultrasonic transmitter and receiver between two feet was developed and validated. The new device consisted of a pressure gauge and a digital centimeter indicator a...
متن کاملEvaluation of human allogeneic collagen gel for correction of nasolabial folds using non-invasive measurement techniques
Background: With aging, the facial folds, grooves, and sagging tissue become more prominent. It is haracterized by loss of the collagen mass in the dermis and an increased array of elastin whirls in the deeper dermis. The aim of this study was to determine whether correction of nasolabial folds could be achieved using an allogeneic collagen product. Methods: Nine healthy volunteers participated...
متن کاملThe Measurement of Rigidity in Elbow Joint: An Objective Method for the Evaluation of Parkinson's disease
Introduction: One of the most common syndromes in Parkinson's disease (PD) is rigidity. Currently, an index is used to evaluate the level of PD by the clinical measurement of rigidity in the upper extremity. The index uses a subjective method called Unified Parkinson's Disease Rating System (UPDRS). The subjective nature of this method makes the influence of physician in the measurement of rigi...
متن کاملIntegrated Technologies Like Noninvasive Brain Stimulation (NIBS) for Stroke Rehabilitation New Hopes for Patients, Neuroscientists, and Clinicians in Iran
A B S T R A C TThe applications of neurophysiological therapy techniques range far and few in the realm of modern day medicine. However, the concept of electromagnetic stimulation, the basis for many noninvasive brain stimulation (NIBS) techniques today, has been of interest to the scientific community since the late nineteenth century. Recently, transcranial direct current stimulation (tDCS) a...
متن کاملClinical Evaluation Tool for Operating Room Students: Development and Measurement of Reliability and Validity
Introduction: In order to ensure the achievement of goals in clinical training, having the right tool with good reliability and validity is essential. Given the fact that no proper tool has ever been developed for evaluation of operating room students, this study aimed to develop and measure the validity and reliability of the evaluation tool for these students. Methods: This descriptive study ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 2 شماره
صفحات -
تاریخ انتشار 2011