Clinical outcome of the PAS-Port® proximal anastomosis system in off-pump coronary artery bypass grafting in 201 patients.
نویسندگان
چکیده
AIM The PAS-Port® Proximal Anastomosis System (Cardica, Inc, Redwood City, CA, USA) has been used worldwide since March 2003. The objective of the present study was to evaluate the clinical outcome of the PAS-Port® Proximal Anastomosis System. METHODS All the patients who underwent off-pump coronary artery bypass grafting in the Catharina Hospital Eindhoven between August 2006 and April 2010 were included in a non-randomized retrospective case-control study, if they had at least one proximal vein graft anastomosis. Study end-points consisted of overall survival, coronary reintervention and postoperative stroke. RESULTS The study included 312 patients (201 cases, 111 controls). After 36 months of follow-up there was no difference in survival between cases and controls (92.2% vs. 93.7%, P=0.52). No significant difference could be detected between cases and controls with respect to overall coronary reintervention-free survival (93% vs. 96.4%, P=0.20) and freedom from coronary reintervention due to proximal vein graft failure (98% vs. 100% P=0.14). The use of the PAS-Port system could not be identified as an independent risk factor of coronary reintervention (p=0.21). Postoperative stroke rates of cases and controls (2% vs. 0.9%, P=0.42) were comparable. CONCLUSION The clinical outcomes in patients treated with the PAS-Port® Proximal Anastomosis System were satisfactory compared with those treated with the conventional hand-sewing technique. The use of the PAS-Port system was not associated with higher adverse outcome in terms of overall survival, stroke, coronary reintervention-free survival and freedom from reintervention due to proximal vein graft failure.
منابع مشابه
Use of skeletonized radial artery graft with the PAS-Port proximal anastomotic device.
PURPOSE We report our initial experience with the PAS-Port proximal anastomosis system (Cardica Inc, Redwood City, CA) using full-skeletonized radial artery (RA) in patients requiring off-pump coronary artery bypass grafting. DESCRIPTION The PAS-Port system (Cardica Inc) was used in 25 patients undergoing off-pump coronary artery bypass surgery. All patients received at least one RA graft usi...
متن کاملA case of ostial stenosis with the PAS-Port proximal anastomosis system in off-pump coronary artery bypass grafting.
The use of an automatic aortic connector device for proximal saphenous vein graft anastomoses eliminates the need for aortic clamping during off-pump coronary artery bypass grafting and may reduce the incidence of stroke in the elderly and in patients with severe aortic atherosclerosis. The PAS-Port proximal anastomotic system is a recently developed sutureless automatic saphenous vein graft an...
متن کامل[Reoperative off-pump coronary revascularization via left thoracotomy using the PAS-Port system to the descending thoracic aorta: report of a case].
We performed redo-off-pump coronary artery bypass grafting( OPCAB) via a left thoracotomy using the PAS-Port system for proximal vein graft anastomoses in a patient with posterolateral myocardial ischemia. The patient was a 76-year-old man who had undergone coronary artery bypass grafting (CABG)[ left internal thoracic artery( LITA)-left anterior descending artery( LAD), saphenous vein graft(SV...
متن کاملStudy of coronary artery bypass using the PAS-Port device: assessment by multidetector computed tomography.
OBJECTIVE The use of automatic anastomotic devices in coronary artery bypass grafting surgery is associated with lower patency rates in comparison to conventional anastomosis methods. This is thought to be caused by graft curvature occurring after closing of the chest wall. METHODS We evaluated 39 grafts in 28 patients who underwent off-pump coronary artery bypass surgery using the PAS-Port. ...
متن کاملAvoidance of Dialysis in an End-Stage Renal Disease Patient Status-post Off-pump Coronary Artery Bypass Grafting
Certain benefits are clearly associated with the use of off-pump coronary artery bypass grafting (CABG) as compared with the on-pump CABG. The superiority is more evident in patients with multiple co morbidities including renal failure.We reviewed the medical records of a 67-year-old male with a past medical history that was significant for multiple cardiovascular diseases and new-onset end-sta...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of cardiovascular surgery
دوره 54 3 شماره
صفحات -
تاریخ انتشار 2013