An objective spinal motion imaging assessment (OSMIA): reliability, accuracy and exposure data
نویسندگان
چکیده
BACKGROUND Minimally-invasive measurement of continuous inter-vertebral motion in clinical settings is difficult to achieve. This paper describes the reliability, validity and radiation exposure levels in a new Objective Spinal Motion Imaging Assessment system (OSMIA) based on low-dose fluoroscopy and image processing. METHODS Fluoroscopic sequences in coronal and sagittal planes were obtained from 2 calibration models using dry lumbar vertebrae, plus the lumbar spines of 30 asymptomatic volunteers. Calibration model 1 (mobile) was screened upright, in 7 inter-vertebral positions. The volunteers and calibration model 2 (fixed) were screened on a motorized table comprising 2 horizontal sections, one of which moved through 80 degrees. Model 2 was screened during motion 5 times and the L2-S1 levels of the volunteers twice. Images were digitised at 5fps. Inter-vertebral motion from model 1 was compared to its pre-settings to investigate accuracy. For volunteers and model 2, the first digitised image in each sequence was marked with templates. Vertebrae were tracked throughout the motion using automated frame-to-frame registration. For each frame, vertebral angles were subtracted giving inter-vertebral motion graphs. Volunteer data were acquired twice on the same day and analysed by two blinded observers. The root-mean-square (RMS) differences between paired data were used as the measure of reliability. RESULTS RMS difference between reference and computed inter-vertebral angles in model 1 was 0.32 degrees for side-bending and 0.52 degrees for flexion-extension. For model 2, X-ray positioning contributed more to the variance of range measurement than did automated registration. For volunteer image sequences, RMS inter-observer variation in intervertebral motion range in the coronal plane was 1.86 degrees and intra-subject biological variation was between 2.75 degrees and 2.91 degrees. RMS inter-observer variation in the sagittal plane was 1.94 degrees. Radiation dosages in each view were below the levels recommended for a plain film. CONCLUSION OSMIA can measure inter-vertebral angular motion patterns in routine clinical settings if modern image intensifier systems are used. It requires skillful radiography to achieve optimal positioning and dose limitation. Reliability in individual subjects can be judged from the variance of their averaged inter-vertebral angles and by observing automated image registration.
منابع مشابه
The Accuracy of Surface Measurement for Osteoporotic Spine Motion Analysis
INTRODUCTION The surface method of measuring lower thoracic and lumbar spine motion with Fastrak ® was examined for validity. There is an increasing awareness of the health risks of exposure to radiation associated with repeated radiographic assessment of spinal curvature and spinal movements. As a skin-surface measurement device, Fastrak ® was employed to study the effect of low bone mineral d...
متن کاملMeasuring Spinal Mobility in Ankylosing Spondylitis: Comparison of Motion Capture Results against Conventional Metrology and Radiological Scores
Ankylosing spondylitis (AS) is a chronic rheumatic disease than affects mainly to the spine reducing its mobility. Spinal mobility restrictions are the measures most commonly used in AS but rheumatologist use basic tools as tapes and goniometers. They also use certain indexes (for example BASMI) that group several measurements. Some studies show lack of reliability, accuracy and responsiveness ...
متن کاملMedical imaging modalities: Prevention of unnecessary orders and non-optimized radiation exposure
Magnetic resonance imaging (MRI) uses a constant magnetic field and radio waves, a non-invasive method for examining tissues, organs and the skeletal system. Advantages and disadvantages of MRI are not fully understood; however, it has been commonly used in the diagnosis of skeleton and bone diseases (such as disk and spinal disorders, joint disorders (arthritis), bone tumours and tissues...
متن کاملReliability of Measuring Lumbar Lordosis, Flexion and Extension Using Dual Inclinometer in Healthy Subjects and Patients with Non-Specific Chronic Low Back Pain
Objective: Accurate assessment of lumbar range of motion is of great value for both evaluating lumbar functions and monitoring treatment progress. Recent research indicates that there is no general consensus on the most valid and reliable method of measuring spinal range of motion. The purpose of this study was to determine the intra-rater reliability of lumbar flexion and extension measurement...
متن کاملQuality assessment of conventional X-ray diagnostic equipment by measuring X-ray exposure and tube output parameters in Great Khorasan Province, Iran
Introduction: Regular implementation of quality control (QC) program in diagnostic X-ray facilities may affect both image quality and patient radiation dose due to the changes in exposure parameters. Therefore, this study aimed to investigate the status of randomly selected conventional radiographic X-ray devices installed in radiology centers of Great Khorasan Province, Iran, to produce the da...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- BMC Musculoskeletal Disorders
دوره 7 شماره
صفحات -
تاریخ انتشار 2006