Controlled clinical comparison of VersaTREK and BacT/ALERT blood culture systems.
نویسندگان
چکیده
To assess the relative yields in automated microbial detection systems of bacteria and yeasts isolated from the blood of adult patients with suspected sepsis, we compared the new VersaTREK system (VTI) (TREK Diagnostic Systems, Cleveland, OH) to the BacT/ALERT 3D system (3D) (bioMérieux, Inc., Durham, NC). Identical protocols were followed for the two systems. Paired aerobic (REDOX 1) and anaerobic (REDOX 2) VTI media were compared with standard aerobic (SA) and anaerobic (SN) 3D media; each of the four culture bottles was filled with 6 to 9 ml of blood. All bottles flagged positive by the instruments were subcultured to determine both true-positive (growth) and false-positive (no growth) cultures. Additionally, to assess false-negative bottles, terminal subcultures were done on all negative companion bottles to true-positive bottles. All isolates were identified by standard methods. All 4 bottles were adequately filled and yielded 413 clinically significant isolates in 5,389 (79%) of the 6,786 4-bottle sets obtained. Although no overall difference in yield or in time to detection was detected between the two systems, significantly more streptococci and enterococci as a group were detected by VTI. Moreover, significantly more microorganisms were detected by VTI for patients receiving antimicrobial therapy. The two systems were comparable (P, not significant) at detecting the 179 unimicrobial episodes of bacteremia seen. False-positive rates for aerobic and anaerobic bottles, respectively, were 1.6% and 0.9% for VTI and 0.7% and 0.8% for 3D. We conclude that the VTI and 3D systems are comparable for detection of bloodstream infections with bacteria or yeasts.
منابع مشابه
Controlled clinical laboratory comparison of two supplemented aerobic and anaerobic media used in automated blood culture systems to detect bloodstream infections.
A 20-ml blood sample was collected from adult patients with suspected bloodstream infections and distributed equally into the four volume-controlled bottles of a blood culture set consisting of aerobic and anaerobic BACTEC Plus/F bottles and aerobic and anaerobic BacT/Alert FAN bottles. All bottles were incubated in their respective instruments for a standard 5-day protocol or until the instrum...
متن کاملControlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems.
In a collaborative study at three university hospitals, the recovery of microorganisms and the speed of detection of microbial growth by the BacT/Alert (Organon Teknika Corporation, Durham, N.C.) and BACTEC 660/730 (Becton-Dickinson Diagnostic Instrument Systems, Sparks, Md.) nonradiometric blood culture systems were compared. A total of 5,918 comparisons were made between BacT/Alert aerobic an...
متن کاملControlled clinical comparison of BacT/alert FA plus and FN plus blood culture media with BacT/alert FA and FN blood culture media.
New blood culture media containing antibiotic-binding polymeric beads have been developed for the BacT/Alert (bioMérieux, Inc., Durham, NC) blood culture system. To assess the performance of these new media, we compared the new BacT/Alert aerobic medium (FA Plus) with resins to BacT/Alert FA medium with activated charcoal and the new BacT/Alert anaerobic medium (FN Plus) to BacT/Alert FN medium...
متن کاملControlled clinical comparison of plastic and glass bottles of BacT/ALERT FA medium for culturing organisms from blood of adult patients.
A new, clear-plastic nonvented aerobic FA bottle, designed to prevent breakage, has been developed for the BacT/ALERT blood culture system. We assessed the new plastic FA bottle by comparing its performance with that of the current glass FA bottle for recovery of microorganisms and time to detection of growth in blood samples obtained for culture from adult patients with suspected bloodstream i...
متن کاملControlled clinical comparison of new pediatric medium with adsorbent polymeric beads (PF Plus) versus charcoal-containing PF medium in the BacT/alert blood culture system.
We conducted a controlled clinical comparison of PF Plus, the new pediatric medium with adsorbent polymeric beads, versus the charcoal-containing PF medium in the BacT/Alert blood culture system. A total of 2,381 pediatric cultures were enrolled, and 1,703 (71.5%) were deemed to be compliant and acceptable for analysis. Seventy-two cultures (4.2%) had a positive result with 80 clinically signif...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of clinical microbiology
دوره 45 2 شماره
صفحات -
تاریخ انتشار 2007