Analysis of the effect of flow rate on the Doppler continuity equation for stenotic orifice area calculations: a numerical study.

نویسندگان

  • C G DeGroff
  • R Shandas
  • L Valdes-Cruz
چکیده

BACKGROUND Flow-rate dependencies of the Doppler continuity equation are addressed in this study. METHODS AND RESULTS By use of computational fluid dynamic (CFD) software with turbulence modeling, three-dimensional axisymmetric models of round stenotic orifices were created. Flow simulations were run for various orifice area sizes (0.785, 1.13, 1.76, and 3.14 cm2) and flow rates (0.37 to 25.0 L/min). Reynolds numbers ranged from 100 to 8000. Once adequate convergence was obtained with each simulation, the location of the vena contracta was determined. For each run, maximum and average velocities across the cross section of the vena contracta were tabulated and vena contracta cross-sectional area (effective orifice area) determined. The difference between the maximum velocity and the average velocity at the vena contracta was smallest at high-flow states, with more of a difference at low-flow states. At lower-flow states, the velocity vector profile at the vena contracta was parabolic, whereas at high-flow states, the profile became more flattened. Also, the effective orifice area (vena contracta cross-sectional area) varied with flow rate. At moderate-flow states, the effective orifice area reached a minimum and expanded at low- and high-flow states, remaining relatively constant at high-flow states. CONCLUSIONS We have shown that significant differences exist between the maximum velocity and the average velocity at the vena contracta at low flow rates. A likely explanation for this is that viscous effects cause lower velocities at the edges of the vena contracta at low flow rates, resulting in a parabolic profile. At higher-flow states, inertial forces overcome viscous drag, causing a flatter profile. Effective orifice area itself varies with flow rate as well, with the smallest areas seen at moderate-flow states. These flow-dependent factors lead to flow rate-dependent errors in the Doppler continuity equation. Our results have strong relevance to clinical measurements of stenotic valve areas by use of the Doppler continuity equation under varying cardiac output conditions.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Doppler Assessment of Prosthetic Valve Orifice Area

Background. Although Doppler echocardiography has been shown to be accurate in assessing stenotic orifice areas in native valves, its accuracy in evaluating the prosthetic valve orifice area remains undetermined. Methods and Results. Doppler-estimated valve areas were studied for their agreement with catheterderived Gorlin effective orifice areas and their flow dependence in five sizes (19/20-2...

متن کامل

Doppler assessment of prosthetic valve orifice area. An in vitro study.

BACKGROUND Although Doppler echocardiography has been shown to be accurate in assessing stenotic orifice areas in native valves, its accuracy in evaluating the prosthetic valve orifice area remains undetermined. METHODS AND RESULTS Doppler-estimated valve areas were studied for their agreement with catheter-derived Gorlin effective orifice areas and their flow dependence in five sizes (19/20-...

متن کامل

Validation of the proximal flow convergence method. Calculation of orifice area in patients with mitral stenosis.

BACKGROUND It has been proposed recently that measuring the flow convergence region proximal to an orifice by Doppler flow mapping can provide a means of calculating regurgitant flow rate. Although verified in experimental models, this approach is difficult to validate clinically because there is no ideal gold standard for regurgitant flows in patients. However, this method also can be used to ...

متن کامل

A new approach for the evaluation of the severity of coarctation of the aorta using Doppler velocity index and effective orifice area: in vitro validation and clinical implications.

Early detection and accurate estimation of COA severity are the most important predictors of successful long-term outcome. However, current clinical parameters used for the evaluation of the severity of COA have several limitations and are flow dependent. The objectives of this study are to evaluate the limitations of current existing parameters for the evaluation of the severity of coarctation...

متن کامل

Effect of six non-Newtonian viscosity models on hemodynamic parameters of pulsatile blood flow in stenosed artery

A numerical study of hemodynamic parameters of pulsatile blood flow is presented in a stenotic artery with A numerical study of hemodynamic parameters of pulsatile blood flow is presented in a stenotic artery with non-Newtonian models using ADINA. Blood flow was considered laminar, and the arterial wall was considered rigid. Studied stenosis severities were 30, 50, and 70% of the cross-section...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 97 16  شماره 

صفحات  -

تاریخ انتشار 1998