Stent coating with titanium-nitride-oxide for reduction of neointimal hyperplasia.

نویسندگان

  • S Windecker
  • I Mayer
  • G De Pasquale
  • W Maier
  • O Dirsch
  • P De Groot
  • Y P Wu
  • G Noll
  • B Leskosek
  • B Meier
  • O M Hess
چکیده

BACKGROUND Coronary stents prevent constrictive arterial remodeling but stimulate neointimal hyperplasia. Stainless steel induces a metallic foreign body reaction, which is absent for titanium. The hypothesis of the present study was that titanium renders the stent surface biologically inert, with reduced platelet and fibrinogen binding. METHODS AND RESULTS Twelve pigs were instrumented with a stainless steel and 2 titanium-nitride-oxide-coated stents (TiNOX 1, ceramic; TiNOX 2, metallic). Animals were restudied after 6 weeks. Histological specimens of stented segments were analyzed by digital morphometry. Platelet adhesion and fibrinogen binding studies were performed in the perfusion chamber. Under in vitro conditions, TiNOX 1 showed reduced platelet adhesion (65+/-3%) compared with TiNOX 2 (72+/-5%; P<0.05) and stainless steel (71+/-4%; P<0.05). Platelet adhesion 48 hours after incubation with human plasma, however, was not different between TiNOX 1 (17+/-3%) and 2 (15+/-3%) but was significantly higher with stainless steel (23+/-2%; P<0.05). Fibrinogen binding was significantly reduced with TiNOX 2 (54+/-3%) compared with TiNOX 1 (82+/-4%, P<0.05) or stainless steel (100%, P<0.05). Histomorphometry revealed a significantly larger neointimal area in stainless steel (2.61+/-1.12 mm(2)) than in TiNOX 1-coated (1.47+/-0.84 mm(2), P<0.02) or TiNOX 2-coated (1.39+/-0.93 mm(2), P<0.02) stents. The reductions were 44% and 47%, respectively. CONCLUSIONS TiNOX coating significantly reduces neointimal hyperplasia in stainless steel stents. The antiproliferative effect was similar for both TiNOX coatings, suggesting that the electrochemical properties are more important for attenuation of neointimal proliferation than the observed differences in platelet adhesion and fibrinogen binding.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of titanium-nitride-oxide-coated stents with zotarolimus-eluting stents for coronary revascularization a randomized controlled trial.

OBJECTIVES This study sought to compare the efficacy of passive stent coating with titanium-nitride-oxide (TiNO) with drug-eluting stents releasing zotarolimus (ZES) (Endeavor, Medtronic, Minneapolis, Minnesota). BACKGROUND Stent coating with TiNO has been shown to reduce restenosis compared with bare-metal stents in experimental and clinical studies. METHODS In an assessor-blind noninferio...

متن کامل

Methotrexate loaded SAE coated coronary stents reduce neointimal hyperplasia in a porcine coronary model.

OBJECTIVE To evaluate the effect of stent based methotrexate delivery on neointimal hyperplasia. METHODS Stainless steel coronary stents and biological polymer coated (SAE) stents were randomly implanted in coronary arteries of pigs with a stent to artery ratio of 1.1:1. The pigs were killed after five days (10 stents) or four weeks (20 stents). Second, stainless steel coronary stents were di...

متن کامل

Drug-eluting stent update 2007: part I. A survey of current and future generation drug-eluting stents: meaningful advances or more of the same?

Analyst projections for the drug-eluting stent (DES) market estimated that the total number of DES implanted in 2010 would go beyond 4.5 million worldwide. Although the initial results seemed promising, longer-term follow-up in a broader range of patients revealed some pitfalls. Delayed neointimal growth, enhanced platelet aggregation, a local hypersensitivity reaction against the polymer coati...

متن کامل

Paclitaxel stent coating inhibits neointimal hyperplasia at 4 weeks in a porcine model of coronary restenosis.

BACKGROUND Despite limiting elastic recoil and late vascular remodeling after angioplasty, coronary stents remain vulnerable to restenosis, caused primarily by neointimal hyperplasia. Paclitaxel, a microtubule-stabilizing drug, has been shown to inhibit vascular smooth muscle cell migration and proliferation contributing to neointimal hyperplasia. We tested whether paclitaxel-coated coronary st...

متن کامل

Part I: A Survey of Current and Future Generation Drug-Eluting Stents: Meaningful Advances or More of the Same?

Analyst projections for the drug-eluting stent (DES) market estimated that the total number of DES implanted in 2010 would go beyond 4.5 million worldwide. Although the initial results seemed promising, longer-term follow-up in a broader range of patients revealed some pitfalls. Delayed neointimal growth, enhanced platelet aggregation, a local hypersensitivity reaction against the polymer coati...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 104 8  شماره 

صفحات  -