Assessing the accuracy and reliability of ultrasonographic three-dimensional parathyroid volume measurement in a patient with secondary hyperparathyroidism: a comparison with the two-dimensional conventional method

نویسندگان

  • Sung-Hye You
  • Sang-Il Suh
  • In-Seon Ryoo
  • Gyu-Ri Son
  • Young-Hen Lee
  • Hyung Suk Seo
  • Nam Joon Lee
  • Hae Young Seol
چکیده

PURPOSE The purpose of this study was to investigate the accuracy and reliability of the semi-automated ultrasonographic volume measurement tool, virtual organ computer-aided analysis (VOCAL), for measuring the volume of parathyroid glands. METHODS Volume measurements for 40 parathyroid glands were performed in patients with secondary hyperparathyroidism caused by chronic renal failure. The volume of the parathyroid glands was measured twice by experienced radiologists by two-dimensional (2D) and three-dimensional (3D) methods using conventional sonograms and the VOCAL with 30°angle increments before parathyroidectomy. The specimen volume was also measured postoperatively. Intraclass correlation coefficients (ICCs) and the absolute percentage error were used for estimating the reproducibility and accuracy of the two different methods. RESULTS The ICC value between two measurements of the 2D method and the 3D method was 0.956 and 0.999, respectively. The mean absolute percentage error of the 2D method and the 3D VOCAL technique was 29.56% and 5.78%, respectively. For accuracy and reliability, the plots of the 3D method showed a more compact distribution than those of the 2D method on the Bland-Altman graph. CONCLUSION The rotational VOCAL method for measuring the parathyroid gland is more accurate and reliable than the conventional 2D measurement. This VOCAL method could be used as a more reliable follow-up imaging modality in a patient with hyperparathyroidism.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Literature review in the treatment of calciphylaxis:A case with uncontrolled and severe secondary hyperparathyroidism

Background: Calciphylaxis is a serious disorder often observed in dialysis patients and less frequently in chronic renal failure patients with secondary hyperparathyroidism. Mortality rate increases following the development of calciphylaxis, immediate application of parathyroidectomy along with other treatment options may be lifesaving. Case Presentation: A 44-year-old male patient had been...

متن کامل

Parathyroid adenoma Localization

 Background: Bilateral neck exploration is the gold standard for parathyroid adenoma localization in primary hyperparathyroidism. But surgeons do not have adequate experience for accurate surgical exploration and new methods are developed for surgery like unilateral exploration and minimally invasive surgery, thus, preoperative localization could reduces time and stress in surgical performance....

متن کامل

Possibility and Reproducibility of Renal Assessing and Size Measurement by Three- Dimensional vs Two- Dimensional Ultrasounographv in Dogs

Objective- To determine the possibility and reproducibility of three-dimensional ultrasonography (3DUS)and comparison of the achieved measurements to normal two-dimensional ultrasonography (2DCS)Design- Descriptive studyAnimals- 10 young mixed normal dogs, age 1.5-2.5 year, weighing 9.7-12 kg Procedure- Renal width, length, and depth measured in coron al and transverse sections. The measurement...

متن کامل

Improvement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method

Background: The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area.Objective: A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of meas...

متن کامل

Ultrasonographic Volumetry of the Canine Thyroid Gland: A Comparison of 2D, 3D Ultrasonography and Real Gland Volume

Objective: to compare the estimated thyroid volume, as determined by two-dimensional (2D) and three-dimensional (3D) ultrasonography, with real volume. Design: Descriptive Study Animals: A total of 7 mixed breed dogs Procedures: On 2D ultrasound, the thyroid volume is estimated by the ellipsoid equation and on 3D ultrasound; it is estimated by outlining the boundaries of the thyroid lobe in ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 36  شماره 

صفحات  -

تاریخ انتشار 2017