Haemolysis following implantation of duct occlusion coils.
نویسندگان
چکیده
OBJECTIVES To describe the incidence and management of haemolysis after transcatheter coil occlusion of the arterial duct. DESIGN Prospective clinical and echocardiographic follow up of patients who have undergone implantation of the Cook detachable duct occlusion coil. SETTING Tertiary paediatric cardiac centre. PATIENTS Five cases of haemolysis (two girls aged 6 and 11 months; three boys aged 6, 17, and 14 months) from a series of 137 duct coil implantations. MAIN OUTCOME MEASURES The occurrence of clinically significant haemolysis after implantation of duct occlusion coils and resolution of haematuria after completion of duct occlusion. RESULTS Haemolysis was detected in five of 137 procedures following implantation of Cook detachable duct coils. Four patients became symptomatic 12 hours after the procedure but in one haemolysis was detected three months later. Resolution of ongoing haemolysis was achieved within 48 hours of detection with further coil implantations, but haematuria persisted for up to 10 days. In one patient the extensive destruction of erythrocytes resulted in acute renal failure requiring peritoneal dialysis. CONCLUSIONS Haemolysis is an important complication after duct coil implantation. It occurred in 3.6% of 137 procedures in this series and is most likely to occur in young patients with relatively large ducts. Further coil implantation to occlude the duct completely is not only successful but technically relatively straightforward and should be undertaken early if major complications such as severe anaemia and renal failure are to be avoided.
منابع مشابه
Early implantation of multiple spring coils for severe haemolysis after incomplete transcatheter occlusion of persistent arterial duct.
An 18 month old girl with an angiographically measured ductus of 4.5 mm underwent transcatheter occlusion of the persistent arterial duct with a 17 mm Rashkind umbrella and an occluding spring coil. Severe intravascular haemolysis developed 20 hours later. Significant residual ductal leakage was noted and the residual duct measured 6 mm. Previous underestimation might have been related to ducta...
متن کاملResidual and recurrent shunts after implantation of Cook detachable duct occlusion coils.
OBJECTIVE To assess the presence and outcome of Doppler detectable shunts following implantation of the Cook detachable PDA coil. DESIGN Prospective study. SETTING Tertiary paediatric cardiac centre. PATIENTS 76 consecutive patients undergoing coil implantation (80 procedures). MAIN OUTCOME MEASURES Detection and colour Doppler echocardiographic appearance of residual or recurrent shunt...
متن کاملEvolving use of embolisation coils for occlusion of the arterial duct.
OBJECTIVE To assess the outcome of arterial duct occlusion with coils chosen according to the duct morphology. DESIGN Retrospective study. SETTING Paediatric cardiology centre. PATIENTS Coil occlusion was attempted in 57 patients aged 0.5 to 15 (median 3.7) years and weighing 5-59 (median 14) kg between January 1991 and December 1995. A residual leak was present in 8 patients after umbrel...
متن کاملLung perfusion studies after detachable coil occlusion of persistent arterial duct.
OBJECTIVE To evaluate relative lung perfusion following complete occlusion of persistent arterial duct with detachable Cook coils. METHODS Ductal occlusion using detachable coils was performed in 35 patients (median age 3.9 years, range 0.5 to 16; 32 native ducts, three patients with previous devices). If the duct could be crossed with a 0.035 inch guidewire and a 4 F catheter after coil impl...
متن کاملSevere haemolysis after percutaneous closure of a ductus arteriosus (arterial duct).
Severe mechanical haemolysis occurred after transcatheter occlusion of a ductus arteriosus (arterial duct) by a Rashkind double umbrella prosthesis. Surgical removal of the device and ligation of the duct were required before haemolysis was abolished.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Heart
دوره 81 2 شماره
صفحات -
تاریخ انتشار 1999