Resistance of humidifiers, and inspiratory work imposed by a ventilator-humidifier circuit.
نویسندگان
چکیده
The pressures and resistances of a bubble humidifier (Bennett Cascade) and a blow-by humidifier (Fisher and Paykel) were measured and computed at gas flow rates from 4.5 to 100 litre min-1. Pressures increased with flows, with the Bennett pressures being greater at all flows. The resistance of the Fisher-Paykel increased with flows, but remained less than that of the Bennett. An inverse resistance-flow relationship was seen with the Bennett up to a flow of 35 litre min-1. The work of breathing through a Servo 900C ventilator-humidifier circuit was computed, using a lung model. Work was performed by the Servo 900C on the lung, especially with the Fisher-Paykel circuit. The Bennett circuit required considerably greater (3.7 times more) inspiratory work. Thus the Bennett Cascade humidifier may present an unacceptable inspiratory load during spontaneous breathing.
منابع مشابه
Comparison of flow-resistive work load due to humidifying devices.
There are many kinds of humidifying devices. We evaluated six humidifiers from the viewpoint of AWLs. The AWL was obtained by calculating the area difference between pressure-volume tracings obtained without and with humidifiers. To examine the effect of pressure monitoring sites on AWL when a humidifier is placed, we measured AWL at three different pressure monitoring sites. The AWL was affect...
متن کاملEquipment review: Mechanical effects of heat-moisture exchangers in ventilated patients
Although they represent a valuable alternative to heated humidifiers, artificial noses have unfavourable mechanical effects. Most important of these is the increase in dead space, with consequent increase in the ventilation requirement. Also, artificial noses increase the inspiratory and expiratory resistance of the apparatus, and may mildly increase intrinsic positive end-expiratory pressure. ...
متن کاملCare of the ventilator circuit and its relation to ventilator-associated pneumonia.
Ventilator circuits should not be changed routinely for infection control purposes. The maximum duration of time that circuits can be used safely is unknown. Evidence is lacking related to ventilator-associated pneumonia (VAP) and issues of heated versus unheated circuits, type of heated humidifier, method for filling the humidifier, and technique for clearing condensate from the ventilator cir...
متن کاملAARC Evidence-Based Clinical Practice Guidelines Care of the Ventilator Circuit and Its Relation to Ventilator-Associated Pneumonia
of Recommendations • Ventilator circuits should not be changed routinely for infection control purposes. The maximum duration of time that circuits can be used safely is unknown. • Evidence is lacking related to ventilator-associated pneumonia (VAP) and issues of heated versus unheated circuits, type of heated humidifier, method for filling the humidifier, and technique for clearing condensate ...
متن کاملMoisturizing and mechanical characteristics of a new counter-flow type heated humidifier.
BACKGROUND During mechanical ventilation effective conditioning of inspired air is important. In this respect, conventional humidifiers do not perform optimally. By design, a counter-flow-type humidifier should improve humidification and heating, but may increase resistance. METHODS We investigated mechanical impedance and work of breathing (using pressure-flow characteristics and additional ...