Mechanically assisted 3D ultrasound guided prostate biopsy system.

نویسندگان

  • Jeffrey Bax
  • Derek Cool
  • Lori Gardi
  • Kerry Knight
  • David Smith
  • Jacques Montreuil
  • Shi Sherebrin
  • Cesare Romagnoli
  • Aaron Fenster
چکیده

There are currently limitations associated with the prostate biopsy procedure, which is the most commonly used method for a definitive diagnosis of prostate cancer. With the use of two-dimensional (2D) transrectal ultrasound (TRUS) for needle-guidance in this procedure, the physician has restricted anatomical reference points for guiding the needle to target sites. Further, any motion of the physician's hand during the procedure may cause the prostate to move or deform to a prohibitive extent. These variations make it difficult to establish a consistent reference frame for guiding a needle. We have developed a 3D navigation system for prostate biopsy, which addresses these shortcomings. This system is composed of a 3D US imaging subsystem and a passive mechanical arm to minimize prostate motion. To validate our prototype, a series of experiments were performed on prostate phantoms. The 3D scan of the string phantom produced minimal geometric distortions, and the geometric error of the 3D imaging subsystem was 0.37 mm. The accuracy of 3D prostate segmentation was determined by comparing the known volume in a certified phantom to a reconstructed volume generated by our system and was shown to estimate the volume with less then 5% error. Biopsy needle guidance accuracy tests in agar prostate phantoms showed that the mean error was 2.1 mm and the 3D location of the biopsy core was recorded with a mean error of 1.8 mm. In this paper, we describe the mechanical design and validation of the prototype system using an in vitro prostate phantom. Preliminary results from an ongoing clinical trial show that prostate motion is small with an in-plane displacement of less than 1 mm during the biopsy procedure.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A molecular image-directed, 3D ultrasound-guided biopsy system for the prostate

Systematic transrectal ultrasound (TRUS)-guided biopsy is the standard method for a definitive diagnosis of prostate cancer. However, this biopsy approach uses two-dimensional (2D) ultrasound images to guide biopsy and can miss up to 30% of prostate cancers. We are developing a molecular image-directed, three-dimensional (3D) ultrasound image-guided biopsy system for improved detection of prost...

متن کامل

A PET/CT Directed, 3D Ultrasound-Guided Biopsy System for Prostate Cancer

Prostate cancer affects 1 in 6 men in the USA. Systematic transrectal ultrasound (TRUS)-guided biopsy is the standard method for a definitive diagnosis of prostate cancer. However, this "blind" biopsy approach can miss at least 20% of prostate cancers. In this study, we are developing a PET/CT directed, 3D ultrasound image-guided biopsy system for improved detection of prostate cancer. In order...

متن کامل

Robotic ultrasound-guided prostate intervention device: system description and results from phantom studies.

BACKGROUND We introduce the first robotic ultrasound-guided prostate intervention device and evaluate its safety, accuracy and repeatability. METHODS The robotic positioning system (RPS) determines a target's x, y and z axes. It is situated with a biplane ultrasound probe on a mobile horizontal platform. The integrated software acquires ultrasound images for three-dimensional (3D) modelling, ...

متن کامل

Closed-Loop Control in Fused MR-TRUS Image-Guided Prostate Biopsy

Multi-modality fusion imaging for targeted prostate biopsy is difficult because of prostate motion during the biopsy procedure. A closed-loop control mechanism is proposed to improve the efficacy and safety of the biopsy procedure, which uses real-time ultrasound and spatial tracking as feedback to adjust the registration between a preoperative 3D image (e.g. MRI) and real-time ultrasound image...

متن کامل

Towards 3D Ultrasound Image Based Soft Tissue Tracking: A Transrectal Ultrasound Prostate Image Alignment System

The emergence of real-time 3D ultrasound (US) makes it possible to consider image-based tracking of subcutaneous soft tissue targets for computer guided diagnosis and therapy. We propose a 3D transrectal US based tracking system for precise prostate biopsy sample localisation. The aim is to improve sample distribution, to enable targeting of unsampled regions for repeated biopsies, and to make ...