Total Body Irradiation Using a Modified Standing Technique: In Vivo Dosimetry with Semiconductor Diodes

نویسندگان

  • Yigit CECEN
  • Bora SINDIR
  • M. Gamze AKSU
  • Beyza S. OZDEMIR
  • Aylin F. KORCUM
  • Nina TUNCEL
  • Melek N. YAVUZ
چکیده

The aim of this study was to establish a fast and simple online dosimetric verification method for total body irradiation (TBI) using semiconductor diodes. Twenty one patients were treated with fractionated TBI (12 Gy, 2 Gy BID) using a modified standing technique between February 2010 and March 2012. The prescribed dose was administered to the patient’s mid-plane at the level of the umbilicus. The lungs were shielded to allow only the absorption of an 8 Gy dose. To verify the dose administered during TBI, 12 semiconductor diodes were attached to the patient’s skin to measure the entrance and exit doses at six anatomical reference points (umbilicus, lungs, neck, forehead, hand, and knee). An adapted version of the arithmetic method which uses a correction factor was used to calculate the mid-plane doses from the entrance and exit dose measurements. The mid-plane doses were then compared with the expected doses. The results of in vivo dosimetry showed that the mid-plane dose calculations conducted using the adapted arithmetic method were consistent with the expectations of the treatment plan. The average percentage dose differences and standard deviations were -0.5 ± 5.3 % for the neck, -2.1 ± 3.7 % for the umbilicus, 2.8 ± 5.2 % for the forehead, and -5.9 ± 11.1 % for the lungs. The extremity doses (hand and knee) were not calculated in the treatment planning program; however they were measured for documentary purposes. In vivo dose verification under TBI conditions was efficiently conducted through the use of semiconductor diodes. The measurement results were consistent with treatment plan dose calculations. The modified standing technique provides a practical method for TBI treatments. The observed heterogeneity is acceptable, the technique does not require additional shields or bolus material, the dosimetric verification is simple and fast, the set-up of the patient is easily reproducible, and the treatment time is within the acceptable limits for a non-dedicated treatment machine.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Achievable accuracy in brain tumors by in vivo dosimetry with diode detectors

ABSTRACT Background: In vivo measurements of applied dose during radiotherapy treatment, is important to ensure accurate dose delivery to patients. Uncertainty in dose delivery should fall within ±5% of the prescribed dose as recommended by ICRU. Assessment of dose for radiotherapy applications performed with various types of detectors. In this study, semiconductor diodes were used which have s...

متن کامل

Verification of the Accuracy of the Delivered Dose in Brain Tumors by in Vivo Dosimetry Using Diode Detectors

Introduction: During radiotherapy, high accuracy in the dose delivery is required because there is a strong  relationship between the absorbed dose, local tumor control and particularly the normal tissue damage. In  many institutions, in vivo dosimetry using diodes is performed to check the actual dose delivered. In general,  the uncertainty in the dose delivered should fall within  ± 5% of the...

متن کامل

Diode calibration for dose determination in total body irradiation

Background: Total Body Irradiation (TBI) is different from standard radiotherapy in many aspects, so it is not easy to predict the delivered dose to the patient under TBI treatment. Diode dosimetry procedures for surface dose reading can help to define variations of the actually delivered dose from the prescribed one. The aim of this study was to describe the measurements made to calib...

متن کامل

In Vivo Dosimetry Using a Flat Surface Sun Nuclear Corporation Diode in 60co Beams for Some Radiotherapy Treatments in Ghana

Introduction: One of the useful standard quality assurance techniques in radiation therapy is monitoring entrance doses in in-vivo dosimetry. An overall tolerance limit of 5% of the absorbed radiation dose has been recommended by the International Commission of Radiological Units. The implementation of an in vivo dosimet...

متن کامل

In vivo dosimetry for total body irradiation: five‐year results and technique comparison

The aim of this work is to establish if the new CT-based total body irradiation (TBI) planning techniques used at University College London Hospital (UCLH) and Royal Free Hospital (RFH) are comparable to the previous technique at the Middlesex Hospital (MXH) by analyzing predicted and measured diode results. TBI aims to deliver a homogeneous dose to the entire body, typically using extended SSD...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2013