The sensitivity of cartilage contact pressures in the knee joint to the size and shape of an anatomically shaped meniscal implant.

نویسندگان

  • M Khoshgoftar
  • A C T Vrancken
  • T G van Tienen
  • P Buma
  • D Janssen
  • N Verdonschot
چکیده

Since meniscal geometry affects the cartilage contact pressures, it is essential to carefully define the geometry of the synthetic meniscal implant that we developed. Recently, six independent modes of size- and shape-related geometry variation were identified through 3D statistical shape modeling (SSM) of the medial meniscus. However, this model did not provide information on the functional importance of these geometry characteristics. Therefore, in this study finite element simulations were performed to determine the influence of anatomically-based meniscal implant size and shape variations on knee cartilage contact pressures. Finite element simulations of the knee joint were performed for a total medial meniscectomy, an allograft, the average implant geometry, six implant sizes and ten shape variations. The geometries of the allograft and all implant variations were based on the meniscus SSM. Cartilage contact pressures and implant tensile strains were evaluated in full extension under 1200N of axial compression. The average implant induced cartilage peak pressures intermediate between the allograft and meniscectomy and also reduced the cartilage area subjected to pressures >5MPa compared to the meniscectomy. The smaller implant sizes resulted in lower cartilage peak pressures and compressive strains than the allograft, yet high implant tensile strains were observed. Shape modes 2, 3 and 6 affected the cartilage contact stresses but to a lesser extent than the size variations. Shape modes 4 and 5 did not result in changes of the cartilage stress levels. The present study indicates that cartilage contact mechanics are more sensitive to implant size than to implant shape. Down-sizing the implant resulted in more favorable contact mechanics, but caused excessive material strains. Further evaluations are necessary to balance cartilage contact pressures and material strains to ensure cartilage protection and longevity of the implant.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Short Term Evaluation of an Anatomically Shaped Polycarbonate Urethane Total Meniscus Replacement in a Goat Model

PURPOSE Since the treatment options for symptomatic total meniscectomy patients are still limited, an anatomically shaped, polycarbonate urethane (PCU), total meniscus replacement was developed. This study evaluates the in vivo performance of the implant in a goat model, with a specific focus on the implant location in the joint, geometrical integrity of the implant and the effect of the implan...

متن کامل

Clinical Results of Meniscal Repair Using Submeniscal Horizontal Sutures

Background:  Parts of the implants placed over the meniscus during meniscal repair can wear down the cartilage in the contact zones and cause chronic synovitis. Placing horizontal sutures under the meniscus may overcome this potential hazard. The purpose of this prospective study was to evaluate the midterm results of arthroscopic meniscal repair using submeniscally placed out-in horizontal sut...

متن کامل

Meniscal Roots: Current Concepts Review

Meniscal root tears are defined as radial tears located within 1 cm from the meniscal attachment or a bony rootavulsion. This injury is biomechanically comparable to a total meniscectomy, leading to compromised hoop stressesresulting in decreased tibiofemoral contact area and increased contact pressures in the involved compartment.These changes are detrimental to the articular cartilage and ult...

متن کامل

Save the Meniscus, A Good Strategy to Preserve the Knee

With increasing life expectancy, the demand for preservation of native articular cartilage is increasing to delay joint arthroplasties. In the knee, there are many different strategies to preserve the articular cartilage including but not limited to corrective osteotomies, chondral injury restoration, intraarticular injection of cells or growth factors(PRP). However, One of the most important s...

متن کامل

The sensitivity of tibiofemoral contact pressure to the size and shape of the lateral and medial menisci.

In an effort to prevent degeneration of articular cartilage associated with meniscectomies, both meniscal allografts and synthetic replacements have been studied. A number of biomechanical criteria may be important for a meniscal replacement to restore normal tibiofemoral contact pressure in the knee joint and hence be clinically successful. One of these criteria is geometric similarity. The ob...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of biomechanics

دوره 48 8  شماره 

صفحات  -

تاریخ انتشار 2015