Electric pumps and hyperbaric treatment.

نویسندگان

  • J P Lévecque
  • I Vincenti-Rouquette
  • J M Rousseau
چکیده

To the Editor:—We read with interest the case report and the study of Sanchez-Guijo et al., which describes the failure of patient-controlled analgesia pumps in a hyperbaric environment. We have experience with this problem. Electronic infusion pumps function in a fashion similar to a patient-controlled analgesia pump. We have observed unexplained low blood pressure in some patients treated by hyperbaric oxygen therapy at 2.5 absolute atmospheres (ATA). These patients were receiving inotropic drugs (epinephrine or norepinephrine), either for therapy of septic shock or for air embolism after cardiac surgery. Alterations in pump (DPS Becton-Dickinson & Company, Franklin Lakes, NJ) performance were suspected. We conducted a series of pressurizations with two pump models: the DPS Becton-Dickinson (the most recent model) and the P 300 (Ivac Medical Systems, Basingstoke, UK). To minimize the risk of explosion, the pumps ran on a battery with less than 50 V, as recommended in French legislation. As in the case report, the DPS pump stopped between 2 and 2.1 ATA. With the P 300 pump, we compared the functioning in a normobaric and a hyperbaric environment at different flows and pressures. In these conditions, the P 300 pump was reliable; the three liquid crystal display screens did not blink, the alarm did not ring, the pump did not stop, and the released flows were equal at each pressure. This difference in reliability could be related to the airtightness of the DPS pump. We believe that pumps could be used after being tested in hyperbaric chambers. There is a real need to establish international norms. Jean-Paul Lévecque, M.D. Staff Anesthesiologist [email protected] Isabelle Vincenti-Rouquette, M.D. Staff Anesthesiologist Fabrice Petitjeans, M.D. Staff Anesthesiologist Jean-Marie Rousseau Professor of Anesthesiology Discipline Anesthésie-réanimation Hôpital d’instruction des Armées du Val de Grâce Paris, France

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance of infusion pumps during hyperbaric conditions.

BACKGROUND Many hyperbaric facilities use infusion pumps inside the chamber. It is therefore important to ensure that this equipment will perform accurately during hyperbaric conditions. The authors tested the function and accuracy of the Imed 965 and Infutec 520 volumetric infusion pumps, the Easy-pump MZ-257 peristaltic infusion pump, and the Graseby 3100 syringe pump. METHODS The authors c...

متن کامل

Comparison of three intravenous infusion pumps for monoplace hyperbaric chambers.

We compared the infusion accuracy of the Baxter Flo-Gard 6201, IVAC 530 and Abbott Lifecare 3HB pumps with saline and enteral formula at chamber pressures from 86.1 kPa (0.85 atm abs) to 304 kPa (3.0 atm abs). The Baxter pump infused +/- 10% saline at all tested pressures and rates (1-1,999 ml/hr). At 1 ml/hour, the IVAC infused 18% more saline than expected (86.1 kPa). The Abbott infused -15% ...

متن کامل

Treatment of nervous system injuries by hyperbaric oxygen: a review of molecular mechanisms

Background and aims: The use of hyperbaric oxygen therapy is one of the treatment methods to treat many diseases and injuries such as decompression sickness, wound healing, carbon monoxide poisoning, thermal burns, and nervous system injuries. This article aims to review the possible mechanisms of the effectiveness of this treatment in neurological injuries according to the available studies...

متن کامل

Glasgow Coma Scale, brain electric activity mapping and Glasgow Outcome Scale after hyperbaric oxygen treatment of severe brain injury.

OBJECTIVE To study the effect of hyperbaric oxy gen (HBO) treatment of severe brain injury. METHODS Fifty-five patients were divided into a treatment group (n=35 receiving HBO therapy) and a control group (n=20 receiving dehydrating, cortical steroid and antibiotic therapy) to observe the alteration of clinic GCS (Glasgow Coma Scale), brain electric activity mapping (BEAM), prognosis and GOS ...

متن کامل

MEDICAL EQUIPMENT FOR MULTIPLACE HYPERBARIC CHAMBERS Part I: Devices for Monitoring and Cardiac Support

Kot J: Medical equipment for multiplace hyperbaric chambers. Part I: Devices for monitoring and cardiac support. Europ J Underwater Hyperbaric Med 2005, 6(4): 115-120. All medical devices introduced into the hyperbaric chamber should be of an appropriate design and fit for use in the hyperbaric environment and they should be certified by the manufacturer for hyperbaric conditions. However, unti...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Anesthesiology

دوره 93 2  شماره 

صفحات  -

تاریخ انتشار 2000